Disinfection:

1. Using the grip clips, place the probe gently into the soak container according to where the cable has been

marked. (Review Probe Marking Instructions on page 6).

CAUTION: Do not let the probe hit the bottom of the container.

CAUTION: Older probes are not water-resistant above the handle and must not be soaked above the
top of the handle. Check with your probe manufacturer to determine whether your probe’s handle or
strain relief can be submerged.

2. Close the door.

3. Soaking time is dependent on your facility’s infection control guidelines. In North America this is between 8

to 20 minutes. (Cidex OPA

®

is 12 minutes)

Rinsing:

1. The initial rinse is done in the rinse container inside the GUS unit. This is an important step to prevent drips

and off-gassing when transferring the probe for the final rinse. After disinfection is complete, push back on
the blue tab and lift the probe up, shake it gently to remove large drops of disinfectant. Then place it directly
into the rinse container while keeping your thumb on the red mark above the tab.

2. Rinse the disinfectant off the probe by

GENTLY

swirling it around in the rinse container. The probe can now

be removed, wiped with a paper towel, and given a final rinse in the sink. This should be for at least 60
seconds under running cold water. Check the disinfectant manufacturer’s instructions for rinsing.

3. Empty the rinse container, flush and then refill with fresh water.

4. For cleaning containers, see

"Cleaning the System."

Repair and Modification

Do not modify or repair the unit without reviewing with CIVCO Medical Solutions; contact CIVCO Medical Solutions
to purchase replacement filters and parts.

Before attempting to perform routine maintenance on the unit, thoroughly review and understand this manual for
complete familiarization with the unit and its operation.

M0046_H

8

G10VPL

GUS Disinfection Soak Station™

Содержание GUS G10VPL

Страница 1: ...741 WWW CIVCO COM Copyright 2018 All Rights Reserved GUS is a registered trademark of CIVCO Medical Solutions All other trademarks are property of their respective owners Printed in the USA G10VPL Wal...

Страница 2: ...Mounting Instructions 4 Using High Level Disinfectants with GUS 6 Step 1 Probe Soaking Instructions 6 Step 2 Filling the Containers 7 Step 3 Ready for Use 7 Repair and Modification 8 Filter Life 9 Fi...

Страница 3: ...ed Safety Precautions Disconnect power before replacing filter or when servicing the unit Refer to the MSDS of the high level disinfectant that you are using for additional safety precautions Environm...

Страница 4: ...inst unsafe practices or potential equipment damage NOTE Indicates additional helpful information Quick Troubleshoot If the GUS system is not running check the following 1 Ensure that the unit is plug...

Страница 5: ...ns This MUST be placed above the workstation l One Bottom support bracket l One filter installed in unit Wall Mounting Instructions 1 You will need the following items before attempting to install thi...

Страница 6: ...p 5 9 Place the unit back over the two top screws Screw the two remaining Philip screws into the bottom bracket and tighten Tighten the top screws CAUTION Do not over tighten the screws 10 For masonry...

Страница 7: ...ble as follows If your probe is NOT water resistant Suspend the probe into an empty container so that the top of the handle is level with the top of the container Depress blue tab and place probe cabl...

Страница 8: ...as there are chemicals inside the containers 5 Fill the remaining container with clean rinse water for the preliminary rinse You MUST have one container for rinse water This ensures that when the pro...

Страница 9: ...complete push back on the blue tab and lift the probe up shake it gently to remove large drops of disinfectant Then place it directly into the rinse container while keeping your thumb on the red mark...

Страница 10: ...ting to the rear of the machine 6 Put the used filter into a plastic bag and dispose of it with other non hazardous waste This does not need to be red tagged 7 Replace the top plate and tighten the ca...

Страница 11: ...spills CIVCO Medical Solutions manufactures Glute Out a glycine based neutralizer part 610 1099 which comes in a convenient single use bottle Glute Out is specifically for use with glutaraldehyde and...

Страница 12: ...BILITY AND OF FITNESS FOR A PARTICULAR PURPOSE ARISING FROM A COURSE OF DEALING OR USAGE ARE SPECIFICALLY EXCLUDED This warranty does not cover conditions and damages resulting from any of the followi...

Отзывы: