Disinfection:
1. Using the grip clips, place the probe gently into the soak container according to where the cable has been
marked. (Review Probe Marking Instructions on page 6).
CAUTION: Do not let the probe hit the bottom of the container.
CAUTION: Older probes are not water-resistant above the handle and must not be soaked above the
top of the handle. Check with your probe manufacturer to determine whether your probe’s handle or
strain relief can be submerged.
2. Close the door.
3. Soaking time is dependent on your facility’s infection control guidelines. In North America this is between 8
to 20 minutes. (Cidex OPA
®
is 12 minutes)
Rinsing:
1. The initial rinse is done in the rinse container inside the GUS unit. This is an important step to prevent drips
and off-gassing when transferring the probe for the final rinse. After disinfection is complete, push back on
the blue tab and lift the probe up, shake it gently to remove large drops of disinfectant. Then place it directly
into the rinse container while keeping your thumb on the red mark above the tab.
2. Rinse the disinfectant off the probe by
GENTLY
swirling it around in the rinse container. The probe can now
be removed, wiped with a paper towel, and given a final rinse in the sink. This should be for at least 60
seconds under running cold water. Check the disinfectant manufacturer’s instructions for rinsing.
3. Empty the rinse container, flush and then refill with fresh water.
4. For cleaning containers, see
"Cleaning the System."
Repair and Modification
Do not modify or repair the unit without reviewing with CIVCO Medical Solutions; contact CIVCO Medical Solutions
to purchase replacement filters and parts.
Before attempting to perform routine maintenance on the unit, thoroughly review and understand this manual for
complete familiarization with the unit and its operation.