background image

 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 
 

 
 
 
 

Model G14SD

 

Vapor Management System 

(For Models: DSD and SSD 110/115v) 

 

 

Instruction Manual 

 

Manufactured in USA for 

Medivators Reprocessing Systems

 by: 

 

CIVCO Medical Solutions 

102 First Street South 

Kalona, IA 52247  USA   

Tel:  

1-800-445-6741 

Fax: 

1-877-329-2482 

Web :   WWW.CIVCO.COM 

 
 

Содержание G14SD

Страница 1: ...em For Models DSD and SSD 110 115v Instruction Manual Manufactured in USA for Medivators Reprocessing Systems by CIVCO Medical Solutions 102 First Street South Kalona IA 52247 USA Tel 1 800 445 6741 Fax 1 877 329 2482 Web WWW CIVCO COM ...

Страница 2: ...ter 1 Functions 4 1 1 General Information 4 1 2 Switch Fuse Location 4 1 3 Specifications 5 Chapter 2 Installation and Setup 6 2 1 The 110 Volt Charcoal Fume Removal System Kit 6 2 2 Installation of adaptor box and G14SD 6 2 3 Installation Instructions 6 Chapter 3 Operation 8 Chapter 4 System Maintenance 8 4 1 Changing the Filters 8 4 2 Cleaning the System 8 Chapter 5 Troubleshooting Guide 9 Limit...

Страница 3: ...ound point Environmental Conditions For indoor use only Altitude up to 2 000 meters Temperature range 5 C to 40 C Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Insulation over voltage category II Pollution degree 2 Supply Ratings 110 120VAC 0 5A 50 60 Hz Fuse 1 0 ...

Страница 4: ...struction manuals in a safe accessible location If you have any questions or comments regarding any information in this instruction manual please contact CIVCO Medical Solutions Repair and Modification This system contains no user serviceable parts Do not modify or attempt to repair it contact CIVCO Medical Solutions to purchase replacement parts Equipment damage can result from user attempts to r...

Страница 5: ...ect to a hospital grade wall mains outlet Fuse Type FSF 5 x 20 mm 1 amp 250 VAC Return vent Filtered air is returned to the room through the bottom vent Ratings plate The model name electrical information and serial number are shown Figure 1 GUS Model G14SD Vapor Management System Installed on Medivators DSD Figure 2 Switch fuse location On Off Switch Ratings Plate and Serial Number Fuse Power Cor...

Страница 6: ...ace Velocity FPM 60 Electrical Voltage 110 120VAC 50 60Hz Current Running 19 watts Current Max 0 5 amps Filters Carbon filter activated carbon Fan Blower backward curved impeller blower Construction Fume Hood powder coated carbon steel ABS plastic Dimensions G14SD Height x Width x Depth 14 3 x 20 5 x 8 Weight 24 lbs ...

Страница 7: ...uctions NOTE 1 These instructions are for mounting the Charcoal Fume Removal System CRFS to the DSD cabinet 2 If the DSD already has the vent louver fitted proceed to step 2 3 8 3 Wear proper protective gear during this procedure 4 If the unit is already hard plumbed call the facility maintenance personnel to disconnect the unit 5 Take extra care while moving the unit if the backside is hard to re...

Страница 8: ...nd make sure the filter is unwrapped and properly in place 2 3 9 Mount the CFRS to the six screw holes at the back of the DSD disinfector using the provided screws MS05 0045 Refer to figure 4 2 3 10 Use the pilot drill to pre drill the bottom screw holes 3 16 inch 4 7mm deep into the back plate of the DSD disinfector Be careful not to drill through the cabinet wall Refer to figure 4 2 3 11 Fasten ...

Страница 9: ...k moisture resistant clothing and chemical resistant gloves Disconnect the system s power supply before removing the filter 1 Turn the system s power switch OFF and disconnect the power cord from the wall outlet 2 Unlatch the top lid 3 Remove the old filter by pulling up on the strap located on top of the filter Place in plastic bag and discard with other non hazardous waste 4 Replace with a new f...

Страница 10: ... Solutions and we will repair or replace the unit at our option during the warranty period CIVCO Medical Solutions assumes no liability for consequential damages of any kind as a result of the use or misuse of the system by the purchaser the purchaser s employees or any others This warranty is expressly in lieu of all other warranties guarantees obligations or other liabilities expressed or implie...

Страница 11: ...el disinfectants for a minimum of six months under normal use If a breakthrough should occur during this period CIVCO Medical Solutions will replace the filter at no charge Registration Card Please complete and return the attached Warranty and Filter Replacement Card to CIVCO Medical Solutions You may complete this warranty card online at WWW CIVCO COM ...

Отзывы:

Похожие инструкции для G14SD