SmartMonitor 2 PS Parents’ Guide
c
omPonent
c
leaning
i
nstruc tions
Soft Carrying Case (optional)
Although the care label in the carrying case suggests machine washing in warm water, the
appearance of the carrying case will change noticeably after washing.
Circadiance
recommends that you wipe the case with a damp cloth or sponge using a light
detergent, if necessary. Air-dry only.
Sensors
Follow the instructions packaged with the sensors for information on cleaning.
P
erforming
a
f
unctional
s
elf
-t
est
The monitor’s
functional self-test checks that all the features of the device are functioning properly. You should
perform a functional self-test at least once a week or according to the instructions given by your
home care
provider. You should also perform the test:
•
After a
lead wire is changed
•
After one of the
patient cables is changed
•
When the monitor has been scanned by airport x-ray machines.
To perform the functional self-test, follow the steps listed below.
•
Insert the ECG patient cable into the socket located on the front of the monitor.
•
Connect the lead wires to the
ECG patient cable. Insert the white lead wire into the opening labeled RA.
Insert the
black lead wire into the opening labeled LA.
•
Connect the lead wires to the functional self-test socket on the
side panel of the monitor. Insert the white
lead wire into the RA opening and then the black lead wire into the LA opening.
•
Insert the oximeter patient cable into the socket located on the front of the monitor.