SmartMonitor 2 PS Parents’ Guide
43
i
f
this
light
is
on
...
a
nd
this
condition
exists
...
i
t
means
...
SpO
2
SpO
2
Continuous red light and
continuous alarm (if enabled)
Check that the SpO
2
probe has not become dislodged from the patient or that the probe has not
been disconnected from the monitor or the oximeter patient cable. If you correct the problem,
the alarm will stop. However, the red light remains on until you press the RESET button.
r
educing
f
alse
a
larms
Proper electrode placement will minimize false alarms.
•
Make sure the electrodes are placed along the mid-line of the side, two finger widths below or lined up
with the nipples.
•
If using the black reusable electrodes with the Velcro belt, ensure the belt is quite snug. Place the electrodes
far enough apart so that when the belt is wrapped around the child, the electrode will be located along the
mid-line of the side, two finger widths below or lined up with the nipples.
•
The skin should be clean and dry; if the skin is unusually dry you may add a few drops of moisture (water)
to the child’s skin prior to
electrode belt placement.
•
When using the black reusable electrodes, make sure that the electrode surface is clean.
•
Use of the third (green - RL) electrode and lead wire is normally not required but may reduce excessive
false low heart rate alarms. Place the green electrode along the outside of the child’s upper thigh.