CI-202 Instruction Manual rev. 6/21/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
Phone (360) 833-8835 Fax (360) 833-1914
12
“Measure”, “scanner”, “auto save”, or “storage” on the right. The bottom line of the display reads
“enter” on the left and the direction arrows on the right.
Setup Measure
<enter>
The options are:
To select another mode using the up or down arrow keys
To select an area to set-up using the right or left arrow keys
To begin the set-up process by pressing the START/ENTER key
Set-Up Measure Parameter
Pressing the START/ENTER key when the “Setup Measure” is displayed on the top line allows
the unit to select between Leaf Area or Root length modes.
Measure = Leaf
<stop><save>
Use the up or down arrow keys to modify the value. Once satisfied with the set-up parameters,
press the SAVE key. To avoid saving the parameters, press the STOP key to abort the process.
Set-Up Scanner Thresholds
Note:
Normally, there is no need to adjust the threshold of the CI-202 because the instruments
are factory-
adjusted to a threshold level that is ideal for “all-around” use.
Manual Threshold Adjustment
First, adjust the threshold with the CI-202 over the reflective tape of the scan board. Select
“Setup Scanner” in the display window and press START. Select “HiThrld” and press START
again. Write down the threshold value displayed. Place a piece of white paper so it covers about
half of the laser scan area starting at the right-hand
side (away from the rail). Select “LoThrld”
and press START. Again, write down the threshold value displayed.
First, calibrate the instrument alone with the instrument over the reflective tape of the scan
board. Select “Setup-Scanner” from the menu, then select “Manual” and press ENTER. The
motor should spin, and the laser should come on. The display should show something like this: