OPERATING INSTRUCTIONS 

MODEL 804R 

DUAL INPUT RTD THERMOMETER 

 
 

 

INTRODUCTION 

This instrument is a 4½ digit, compact-sized portable 

digital thermometer designed to use external 100

 Plati-

num RTD as temperature sensor. Temperature indication 
follows Reference Temperature/Resistance Tables (Pt385 
for European Curve, Alpha=.00385. Pt3926 for American 
Curve, Alpha=.003926.) 
Features of the instrument include: 

y

 Dual Input Function 

y

 Pt3926/Pt385 RTD Standard   

y

 According to temperature standard ITS-90 

y

 Resistance Measurement 

y

 One-second measurement cycle 

y

 Real Time indicate 

y

 MAX/MIN with elapsed time 

y

 AVG, REL, HOLD function 

y

 Warning beeper with Hi/Lo setting 

y

 Time setting 

y

 REL Setting 

y

 °C/°F user-selective 

y

 Temperature resolution: 0.1°C or 0.1°F 

y

 Reading Resistance resolution: 0.1

 

y

 Reading Resistance resolution by PC: 0.01

 

y

 Multiple signal: 3000 samples with start-time data 

y

 Single signal: 32 samples with real-time data 

y

 Support user-linear table setting by PC 

y

 USB Mini-B interface standard 

y

 Optional 12V DC power 

 

SAFETY INFORMATION 

It is recommended that you read the safety and opera-

tion instructions before using the thermometer. 

WARNING 

To avoid electrical shock, do not use this instrument 

when working voltages at the measurement surface 

over 24V AC or DC.

 

 

WARNING 

To avoid damage or burns, do not make temperature 

measurement in microwave ovens.

 

 

SPECIFICATIONS 

ELECTRICAL 

Temperature Scale: 

Celsius or Fahrenheit user-selectable. 

Measurement Range: 

Pt385(100

) -200°C to 800°C, (-328°F to 1472°F) 

Pt3926(100

) -200°C to 630°C, (-328°F to 1166°F) 

Temperature Resolution: 

0.1°C or 0.1°F   

Temperature Accuracy:   

±(0.05% rdg + 0.2°C) on °C scale 
±(0.05% rdg + 0.4°F) on °F scale 

Accuracy is specified for operating temperatures over the 
range of 18°C to 28°C (64°F to 82°F), for 1 year, not in-
cluding RTD probe error. 

Temperature Coefficient: 

0.1 times the applicable accu-

racy specification per °C from 0°C to 18°C and 28°C 
to 50°C  (32°F to 64°F and 82°F to 122°F). 

Resistance Range: 

0 to 400

 

Resistance Resolution:

   

0.1

(Meter Reading) 

0.01

(PC Reading) 

Resistance Accuracy: 

±(0.05% rdg + 2dgts) Meter Reading 
±(0.05% rdg + 4dgts) PC Reading 

Input Protection: 

24V dc or 24V ac rms maximum input   

voltage on any combination of input pins.

 

Input Connector: 

Accepts for RTD 4 wires subminiature   

4-Prong type connectors. 

Maximum Differential Common Mode Voltage: 

1volt. (Maximum Voltage between T1 and T2 during   
measurement) 

 
Back Side:

 

1. USB Port 
2. DC power JACK(12V) 
 
 

ENVIRONMENTAL 

Ambient Operating Ranges: 

0°C to 50°C (32°F to 122°F) <70% R.H. 

Storage Temperature: 

-20°C to 60°C (-4°F to 140°F) <80% R.H. 

 

GENERAL 

Display: 

4½ digit liquid crystal display (LCD) with 

maximum reading of 19999. 

Overload:

 “----.-”or “OL” is display. 

Battery:

 1.5V x 4 PCS (SIZE AAA) UM-4 R03. 

Battery Life: 

120 hours typical with carbon zinc battery. 

Reading Rate: 

1 times per second. 

Auto Power Off:

 30 minutes, press power key to resume 

operation. 

Dimensions:

 160mm(H) x 83mm(W) x 38mm(D).

 

Weight: 

Approx. 260g including batteries. 

 

OPERATING INSTRUCTIONS 

1. “ ”

 

Power Switch 

The “ ” key turns the thermometer on or off. In the 

SET mode can not be powered off. Exit SET mode to 
power off. 

    APO function mode 

Press “

” power key for more than 6 seconds to 

disable the auto-power function. The display will show 
“APO OFF”. 

2. “°C/°F/

” Selecting Button 

To change the scales of °C/°F/

, press the “°C/°F” 

key to cycle through °C, °F and 

. When the ther-

mometer is turned on, it is set to the temperature scale 
that was in use when the thermometer was last turned 
off. 

3. “ ” Display Back-Light 

Press the “

” key to turn on or turn off the 

Back-Light. 

4. “T1 T2/T1-T2” Main display Input Selection 

The input selection indicates which input is selected 

for main display; T1 probe, T2 probe or the difference 
between the two probes (T1-T2). When the thermometer 
is turned on, it is set to T1, when main display input se-
lected T1, then T1 input can select alternate of probe by 
pressing “Pt385/Pt3926” key switch. 

5. “TYPE” (Pt385/Pt3926)Main display Input 

RTD Probe Select 

The “TYPE” key switches the T1 input to select the 

Pt385 or Pt3926 or User Liner RTD probe as input 
when main display input selected T1. It is invalid for the 
selection of these parameters when the main display se-
lects T1-T2. When the thermometer is turned on, it is set 
to the probe selected that was in use when the ther-
mometer was last turned off. Switch flow: Pt385, 
Pt3926, User Linear. 

(*)User Linear please refer section 16. 

 6. “MAX/MIN” 

With Time record Mode (only Main display) 

Press “MAX/MIN” key to enter the MIN MAX Re-

cording mode, (displays the Maximum reading with 
time, Minimum reading with time and Average reading 
stored in recording mode). In the this mode the auto-
matic power-off feature is disabled and “

” key, 

“°C/°F” key, “REL” key, “SET” key, “Hi/Lo LIMITS” 
key and main display “T1 T2 T1- T2” key, “TYPE” key 
are disabled. The beeper emits a tone when a new 
minimum or maximum value is recorded. 

Push “MAX/MIN” key to cycle through the MAX, 

MIN and AVG readings. 

If an overload is recorded, the averaging function is 

stopped. In this mode, press “HOLD” key to stop the 
recording of readings, all values are frozen, press again 
to restart recording. To prevent accidental loss of MAX, 
MIN, and AVG data, this mode can only be cancelled 
by pressing and holding the “MAX/MIN” key for 2 
seconds. All recorded readings are erased. 

  7. “REL” Relative Mode(only Main display) 

Press the “REL” key to enter the relative mode, zero 

the display, and store the displayed Reading as a refer-
ence value. REL is shown on the display. Press “REL” 
key again to exit the relative mode. The relative refer-
ence value can also be entered by the user. (See “SET 
mode” later in this manual). When the desired relative 
value has been entered, press “REL” key to enter the 
relative mode and than press “SET” key use the entered 
relative value as a reference value. Press “REL” key 
again to exit the relative mode. In the relative mode, the 
value (can not >±1999.9 counts) shown on the LCD is 
always the difference between the stored reference and 
the present reading. 

8. “HOLD” Mode(only Main display) 

Press the “HOLD” key to enter the Data Hold mode, 

the “HOLD” annunciator is displayed. When HOLD 
mode is selected, the thermometer held the present 