CHS MED-RX Series Скачать руководство пользователя страница 1

 

 

Instruments

 

 

Instructions

 

for

 

Use

 

Distributed in Canada by: 

 Rev.IN-INSTRUM.0 

 

Canadian

 

Hospital

 

Specialties

 

Ltd.

  

Canadian

 

Hospital

 

Specialties

 

|

 

2810

 

Coventry

 

Road,

 

Oakville,

 

Ontario

 

L6H

 

6R1

  

1

800

461

1423

 

|

 

[email protected]

 

|

 

www.chsltd.com

CLEANING

 

AND

 

MAINTENANCE

 

A)

 

Rinsing

 

Directly

 

after

 

surgery,

 

rinse

 

instruments

 

under

 

hot

running

 

water,

 

removing

 

all

 

body

 

fluids

 

and

 

debris.

 

B)

 

Disinfecting

 

(the

 

protection

 

for

 

medical

 

personnel

 

from

 

accidental

 

contamination

 

during

 

cleaning)

 

To

 

avoid

 

blood

 

and

 

other

 

proteins

 

from

 

sticking

 

to

 

instrument

 

surfaces,

 

an

 

enzymatic

 

cleaner

 

bath

 

(soaking)

 

should

 

be

 

used

 

on

 

all

 

instruments.

 

After

 

soaking

 

for

 

a

 

minimum

 

of

 

10

 

minutes,

 

rinse

 

all

 

instruments

 

in

 

running

 

tap

 

water.

 

Immerse

 

instruments

 

completely

 

in

 

hospital

 

approved

 

disinfectant

 

for

 

an

 

additional

 

10

 

minutes

 

then

 

rinse

 

again

 

under

 

running

 

water.

 

DO

 

NOT

 

USE

 

BLEACH.

 

Exposure

 

to

 

bleach

 

will

 

damage

 

instruments.

 

C) Cleaning

 

Instruments

 

should

 

be

 

submerged

 

in

 

a

 

solution

 

of

 

water

 

and

 

neutral

 

pH

 

(7)

 

detergent.

Ultrasonic

 

Cleaning

 

 

Place

 

instruments

 

in

 

an

 

open

 

position

 

into

 

the

 

ultrasonic

 

cleaner.

 

Make

 

sure

 

all

 

“sharp”

 

(i.e.

 

scissors,

 

knives,

 

blades,

 

etc.)

 

do

 

not

 

touch

 

other

 

instruments.

 

All

 

instruments

 

must

 

be

 

completely

 

submerged.

 

Instruments

 

should

 

be

 

processed

 

in

 

the

 

ultrasonic

 

cleaner

 

for

 

the

 

full

 

recommended

 

cycle

 

time,

 

which

 

is

 

typically

 

5

10

 

minutes.

 

A

 

lid

 

should

 

cover

 

the

 

ultrasonic

 

cleaner

 

during

 

the

 

operation

 

to

 

avoid

 

splashing.

 

Never

 

place

 

different

 

metals

 

in

 

the

 

same

 

cleaning

 

cycle.

 

For

 

example,

 

stainless

 

steel

 

and

 

copper,

 

etc.

  

Change

 

solution

 

frequently

 

per

 

manufacturer

 

recommendations.

 

Rinse

 

all

 

instruments

 

after

 

an

 

ultrasonic

 

cleaning

 

with

 

water

 

to

 

remove

 

the

 

excess

 

cleaning

 

solution.

 

Automatic

 

Washing

 

Sterilization

 

Follow

 

all

 

manufacturers’

 

recommendations.

  

Ensure

 

all

 

instruments

 

are

 

lubricated

 

after

 

and

 

before

 

the

 

sterilization

 

cycle.

 

Manual

 

Cleaning

 

Ultrasonic

 

cleaning

 

is

 

the

 

preferred

 

method

 

of

 

cleaning

 

and

 

strongly

 

recommended

 

to

 

ensure

 

all

 

dirt,

 

debris

 

and

 

are

 

fully

 

removed

 

 

particularly

 

with

 

instruments

 

with

 

hinges,

 

locks,

 

and

 

other

 

moving

 

parts.

   

If

 

ultrasonic

 

cleaning

 

is

 

not

 

available,

 

follow

 

the

 

below

 

instructions.

 

a)

 

Use

 

stiff

 

nylon

 

cleaning

 

brushes.

   

Do

 

not

 

use

 

steel

wool

 

or

 

wire

 

brushes

 

unless

 

specifically

 

recommended

 

for

 

cleaning

 

brushes

 

for

 

instrument

serrated

 

areas

 

or

 

on

 

bone

 

files,

 

burs

 

or

 

on

 

stained

 

areas

 

in

 

knurled

 

handles.

Recommended

 

Cleaning

 

Procedure

 

for

 

Cleaning

 

Maintenance

 

and

 

Sterilization

 

of

 

Surgical

 

Instruments

 

b) Use

 

only

 

neutral

 

pH

 

(7)

 

detergents,

 

low

 

pH

 

(acidic

 

less

 

than

 

6

 

pH)

 

detergents

 

will

 

cause

 

breakdown

 

of

 

stainless

 

steel

 

protective

 

surface

 

(pitting)

 

and

 

black

 

staining.

Higher

 

pH

 

detergents

 

(alkaline

 

 

more

 

than

 

8

 

pH)

 

will

cause

 

a

 

surface

 

deposit

 

of

 

a

 

brown

 

stain

 

(phosphates),

 

which

 

will

 

also

 

interfere

 

with

 

a

 

smooth

 

operation

 

of

 

the

 

instrument.

 

Most

 

brown

 

stains

 

are

 

not

 

rust,

 

but

 

merely

 

a

 

high

 

pH

 

surface

 

(phosphate)

 

deposit

 

and

 

can

 

easily

 

be

 

removed

 

with

 

a

 

stain

 

remover.

c) Brush

 

delicate

 

instruments

 

carefully.

  

These

 

should

 

be

handled

 

separately

 

from

 

all

 

other

 

general

 

instruments.

 

d) Make

 

sure

 

all

 

instrument

 

surfaces

 

are

 

visibly

 

clean

 

and

 

free

 

from

 

stains

 

and

 

tissue.

 

The

 

stain

 

remover

 

will

 

aid

 

in

 

the

 

elimination

 

of

 

residue

 

stains.

Each

 

instrument

 

should

 

be

 

inspected

 

for

 

functionality

 

and

 

general

 

condition,

 

this

 

would

 

include,

 

but

 

is

 

not

 

limited

 

to:

 

Forceps

 

have

 

properly

 

aligned

 

tips.

Hemostats

 

and

 

needle

 

holders

 

should

 

not

 

show

 

light

 

between

 

the

 

jaws,

 

when

 

closed

 

in

 

the

 

first

 

ratchet

 

position

 

(hemostats

 

may

 

show

 

a

 

small

 

open

 

space

 

halfway

 

in

 

from

 

the

 

closed

 

tips),

 

lock

 

and

 

unlock

 

easily

 

at

 

joints

 

and

 

make

 

sure

 

they

 

are

 

not

 

too

 

loose.

  

Check

needle

 

holders

 

for

 

wear

 

on

 

jaw

 

surfaces.

Check

 

suction

 

tubes

 

to

 

ensure

 

they

 

are

 

clean

 

inside.

Ensure

 

biopsy

 

punches

 

are

 

clean

 

by

 

punching

 

a

 

hole

 

into

 

tissue

 

paper.

 

Check

 

working

 

functions

 

of

 

all

 

retractors.

 

Make

 

sure

 

that

 

all

 

cutting

 

instruments

 

and

 

knives

have

 

sharp

 

undamaged

 

blades.

e)

 

After

 

scrubbing,

 

rinse

 

all

 

instruments

 

thoroughly

 

under

 

running

 

water.

 

While

 

rinsing,

 

open

 

and

 

close

 

scissors,

 

hemostats,

 

needle

 

holders

 

and

 

other

 

hinged

 

instruments

 

to

 

make

 

sure

 

the

 

hinged

 

areas

 

are

 

rinsed

 

out,

 

as

 

well

 

as

 

the

 

outside

 

of

 

the

 

instruments.

D)

 

After

 

Cleaning

 

If

 

instruments

 

are

 

to

 

be

 

stored,

 

let

 

them

 

air

 

dry

 

then

 

store

 

them

 

in

 

a

 

clean

 

and

 

dry

 

area.

E)

 

Autoclaving

 

Lubricate

 

all

 

instruments

 

that

 

have

 

any

 

“metal

 

to

 

metal”

action

 

(i.e.

 

scissors,

 

hemostats,

 

etc.).

 

Non

silicone,

 

water

soluble

 

surgical

 

lubricants,

 

such

 

as

 

spray

 

lube

 

is

recommended.

 

Do

 

not

 

sure

 

industrial

 

lubricants.

 

Put

 

instruments

 

up

 

for

 

sterilization

 

either

 

individually

 

or

 

in

 

sets.

For

 

Individual

 

Instruments

 

 

Disposable

 

paper

 

or

 

plastic

 

pouches

 

are

 

ideal.

 

Make

 

sure

 

you

 

use

 

a

 

wide

 

enough

 

pouch

 

(4

 

inches

 

or

 

wider)

 

for

 

instruments

 

with

 

ratchet

 

locks

 

(i.e.

 

hemostats

 

and

 

needle

 

holders,

 

etc.)

 

so

 

the

 

instrument

 

can

 

be

 

sterilized

 

in

 

an

 

open

 

(unlocked)

 

position.

 

Locking

 

instruments

 

during

 

autoclaving

 

will

 

result

 

in

 

cracked

 

hinges

 

and

 

other

 

defects

 

because

 

of

 

heat

 

expansion.

 

If

 

you

 

wrap

 

instruments,

 

make

 

sure

 

your

 

towels

 

do

 

not

 

contain

 

detergent

 

residue,

 

which

 

can

 

stain

 

your

 

instruments.

 

Make

 

sure

 

that

 

the

 

towels

 

used

 

in

 

sterilization

 

of

 

instruments

 

have

 

no

 

detergent

 

residue

 

and

 

are

 

neutral

 

pH

 

(7)

 

if

 

immersed

 

in

 

water.

 

This

 

can

 

be

 

a

 

problem,

 

as

 

laundries

 

will

 

frequently

 

use

 

inexpensive

 

but

 

high

 

pH

 

(9

13)

 

detergents

 

and

 

do

 

not

 

properly

 

rinse

 

out

 

or

neutralize

 

those

 

detergents

 

in

 

the

 

final

 

wash

 

or

 

rinse

 

cycle.

 

Also,

 

sometimes

 

bleaches

 

are

 

added

 

and

 

are

 

not

neutralized.

 

Hospitals

 

use

 

a

 

“sour”

 

rinse

 

cycle

 

to

 

neutralize

 

all

 

detergent

 

residue.

 

For

 

Instrument

 

Sets

 

 

Unlock

 

all

 

instruments

 

and

 

sterilize

 

them

 

in

 

an

 

open

 

position.

 

Place

 

the

 

heavier

 

instruments

 

on

 

the

 

bottom

 

of

 

the

 

set,

 

when

 

two

 

layers

 

are

 

required.

 

Never

 

lock

 

an

 

instrument

 

during

 

autoclaving.

 

Steam

 

cannot

 

penetrate

 

contacting

 

surfaces

 

and

 

this

 

will

 

compromise

 

the

 

sterility

 

of

 

the

 

instruments.

 

The

 

instrument

 

will

 

develop

 

cracks

 

in

 

hinges

 

(lock

 

box)

 

areas

 

because

 

of

 

the

 

heat

 

expansion

 

during

 

sterilization.

 

Do

 

not

 

overload

 

autoclave

 

chambers

 

as

 

pockets

 

may

 

form

 

and

 

will

 

not

 

permit

 

steam

 

to

 

penetrate.

 

Place

 

a

 

towel

 

on

 

the

 

bottom

 

of

 

the

 

pan

 

to

 

absorb

 

the

 

excess

 

moisture

 

during

 

autoclaving.

 

CAUTION

:

 

With

 

most

 

portable,

 

tabletop

 

autoclaves,

 

at

 

the

 

end

 

of

 

the

 

autoclave

 

cycle

 

 

before

 

drying

 

cycle

 

 

unlock

 

the

 

door

 

and

 

open

 

it

 

no

 

more

 

than

 

a

 

crack

 

(approximately

 

1/4”).

 

Then

 

run

 

dry

 

cycle

 

for

 

the

 

period

 

recommended

 

by

 

the

 

autoclave

 

manufacturer.

 

If

 

the

 

autoclave

 

door

 

is

 

opened

 

fully

 

before

 

the

 

drying

 

cycle,

 

cold

 

room

 

air

 

will

 

rush

 

into

 

the

 

chamber,

 

causing

 

condensation

 

on

 

the

 

instruments.

 

This

 

will

 

result

 

in

 

water

 

stains

 

on

 

the

 

instruments

 

and

 

also

 

cause

 

wet

 

packs.

 

Make

 

sure

 

the

 

autoclave

 

filters

 

and

 

chambers

 

are

 

cleaned

 

periodically.

 

A

 

stain

 

remover

 

will

 

remove

 

stains

 

and

 

clean

 

the

 

autoclave

 

chamber.

 

F) Cold

 

Sterilization

 

Most

 

cold

 

sterilization

 

solutions

 

render

 

instruments

 

sterile

 

only

 

after

 

10

hour

 

immersions.

 

This

 

prolonged

 

chemical

 

action

 

can

 

be

 

more

 

detrimental

 

to

 

surgical

 

instruments

 

than

 

the

 

usual

 

20

minute

 

autoclave

 

cycle.

 

If

 

the

 

instrument

 

needs

 

to

 

be,

 

“disinfected”

 

only,

 

a

 

cold

 

sterilization

 

soak

 

is

 

okay,

 

as

 

disinfecting

 

will

 

take

 

place

 

in

 

10

minutes

 

or

 

more.

 

Check

 

the

 

manufacturer’s

 

specifications.

 

Also,

 

see

 

the

 

warning

 

sign

 

on

 

using

 

bleach

 

(section

 

B).

 

Keep

 

in

 

mind

 

the

 

difference

 

between:

  

STERILE

 

 

No

 

living

 

microorganisms

   

DISINFECTED

 

 

Kills

 

most

 

but

 

not

 

all

 

microorganisms

 

CAUTION

:

 

For

 

instruments

 

with

 

Tungsten

 

Carbide

 

inserts

 

(hemostats,

 

scissors,

 

needle

 

holders

 

 

also

 

identified

 

by

 

a

 

gold

 

handle),

 

we

 

do

 

not

 

recommend

 

cold

 

sterilization

 

or

 

solutions

 

containing

 

Benzyl

 

Ammonium

 

Chloride,

 

which

 

will

 

deteriorate

 

the

 

Tungsten

 

Carbide,

 

inserts.

 

*French instructions to follow/* Mode d’emploi en français à suivre

Отзывы: