22
GENERAL INFORMATION
GENERAL SPECIFICATIONS
Case
:
ABS grey color (RAL 7043); self-extinguish-
ing degree: V-0 according to UL 94.
Front protection
- designed and tested for IP 65 (*)
and NEMA 4X (*) for indoor locations (when panel
gasket is installed).
(*) Test were performed in accordance with CEI 70-
1 and NEMA 250-1991 STD.
Installation
: panel mounting by means of mounting
braket. Instrument removable from case.
Rear terminal block
:10 screw terminals ( screw
M3, for cables from
f
0.25 to
f
2.5 mm
2
or from
AWG 22 to AWG 14 ) with connection diagrams
and safety rear cover.
Dimensions
: 48 x 48 mm, depth 100 mm (DIN
43700).
Weight
: 160 g max.
Power supply
:
- 100V to 240V AC 50/60Hz (-15% to + 10% of the
nominal value).
- 24 V AC/DC (+ 10 % of the nominal value).
Power consumption
: 6 VA max.
I
nsulation voltage
: 2300 V rms according to
EN 61010-1.
Display updating time
: 500 ms.
Sampling time:
500 ms.
Resolution
: 30000 counts.
Accuracy
(@ 25 °C): +0.3% of the input span +1 °C.
Common mode rejection
: 120 dB at 50/60 Hz.
Normal mode rejection:
60 dB at 50/60 Hz.
Electromagnetic compatibility and safety
requirements
: This instrument is marked CE.
Therefore, it is conforming to council directives
89/336/EEC (reference harmonized standard
EN-50081-2 and EN-50082-2) and to council
directives 73/23/EEC and 93/68/EEC (reference
harmonized standard EN 61010-1).
Installation category
: II
Temperature drift
: < 200 ppm/°C (RJ excluded)
< 400 ppm/°C for RTD input with -19.9/99.9 °C range
and TC type T.
Reference junction drift
: 0.1 °C/°C.
Operative temperature
: from 0 to 50 °C (32 to
122°F).
Storage temperature :
-20 to +70 °C (-4 to 158°F).
Humidity
: from 20 % to 85% RH, non condensing.
Protections
:
1) WATCH DOG circuit for automatic restart.
2) DIP SWITCH for protection against tampering
of configuration and calibration parameters.
INPUTS
A) THERMOCOUPLE
Type
: L, J, K, N, T programmable by front
pushbuttons.
Line resistance
: max. 100
W
with error <+0.1% of
the input span.
Engineering unit
: °C or °F programmable.
Reference junction
: automatic compensation
from 0 to +50 °C.
Reference junction drift
: 0.1 °C/°C.
Burn-out
: Up or down scale selectable.
Calibration
: according to IEC 584-1 and DIN
43710 - 1977 (TC L)
STANDARD RANGES TABLE
TC
type
L
J
K
N
T
0 - +800 °C
0 - +800 °C
0 - +999 °C
0 - +999 °C
0 - + 400 °C
Measuring ranges
0 - +999 °F
0 - +999 °F
0 - +999 °F
0 - +999 °F
0 - +752 °F
1601-1-0C.p65
5/2/00, 1:45 PM