Power Supply Operation
3-13
signal will show at the bottom right like the picture “Slave State.” However, when the master
state is enabled, there is no signal shown at the bottom right.
1. Press the
Menu
button.
2. Rotate the Knob to show MA/SL on the display.
3. Press the
Menu
button to confirm.
4. Rotate the Knob to show STATE on the display.
5. Press the
Menu
button to confirm.
6. Rotate the Knob to show ON on the display.
7. Press the
Menu
button to finish the settings.
To quit the configuration, press any other key to exit.
When the above settings are done, but before enabling the output on the master unit, the
display on the master unit should display S-MA OFF and slave unit should show S-SL OFF as
in the picture to the right side.
Unit 1, set the slave condition to the output off state.
Unit 2, set the slave condition to the output off state.
Unit 3, set the slave condition to the output off state.
After enabling the output on the master unit, the display on the master unit shows a voltage
value and a current value. On the display of the salve units, only a voltage value will be
shown.
Unit 1, set the master condition to the output on state.
Unit 2, set the slave condition to the output on state.
Unit 3, set the slave condition to the output on state.
3.8.2 Connections in Parallel
To output current within the 0A to 49A range, parallel operation of two or more power supplies
is necessary. Using parallel connections, CV/CC capability, and the automatic crossover
operation, you can obtain a total output current greater than that available from single power
supply. The total output current is the sum of the currents from the connected power supplies.
The output of each power supply can be set separately. The output voltage controls of the
Master power supply should be set to the desired output voltage. The Slaves will work in CC
mode with the CC Limit the same as the master output current.
When power supplies operate in parallel, there must less than three of
power supplies can enable remote sensing function.
CAUTION