19501 Series Fixture B195000/B195001/B195002 Quick Start Guide
14
5.2
Fixture Function and Installation
The functions of the B195002 front terminals are shown in the figure
below.
(1)
RTN/Low
: This terminal connects to the RTN/LOW
terminal on the high voltage module of the
19501 partial discharge tester, which is a low
potential terminal. This terminal is almost
equal to the ground terminal on the chassis.
(2)
Low Contact
: This terminal connects to the HVCC-RTN
terminal on the high voltage module of the
19501 partial discharge tester.
(3)
HV Contact
: This terminal connects to the HVCC-HV
terminal on the high voltage module of the
19501 partial discharge tester.
(4)
HV
: This terminal connects to the HV terminal on
the high voltage module of the 19501 partial
discharge tester. Do not touch during testing
and when the DANGER light is on.
(5)
Micro switch
: This switch connects to the Interlock terminal
on the fixture rear side. When the acrylic