3. Open the shutter.
4. From the Test Pattern menu, select the Boresight test pattern.
The Boresight test pattern assists with adjusting the boresight for the three focus points.
5. To focus the right boresight guide, adjust the green boresight screw.
6. To focus the left boresight guide, adjust the yellow boresight screw.
7. To continue to refine the focus, repeat steps 5 to 7.
8. Before adjusting the stabilization screws, close the shutter.
9. To maintain the adjustments, lock the stabilization screws.
When locking the screws, start with the top stabilization screw and turn it so it just touches the
base. Repeat for the other two stabilization screws. Continue to adjust the locking screws until
they are tight.
10. Open the shutter.
Optimizing the integrator zoom and focus
Learn how to optimize the integrator zoom and focus.
1. Set the brightness to 30% or less before powering on the projector.
If the Brightness slider is greyed out, enable LiteLOC
™
.
2. From the Test Pattern menu, select the Integrator Rod test pattern.
Installation and setup
Christie Griffyn Series Installation and Setup Guide
38
020-103314-08 Rev. 1 (06-2022)
Copyright
©
2022 Christie Digital Systems USA, Inc. All rights reserved.
If printed, please recycle this document after use.
Содержание Griffyn 4K50-RGB
Страница 1: ...Installation and Setup Guide 020 103314 08 Christie Griffyn Series...
Страница 86: ......