   D4K and Mirage 4K Troubleshooting

D4K and Mirage 4K Troubleshooting Technical Reference

6

020-101452-03 Rev. 1 (10-2014)

Details

In a multi-port input configuration, the projector requires all signals to have a stable sync. The 
projector automatically selects the top-left signal as the master signal to use as reference. If the 
master is unstable, all of the other signals look unstable. Possible symptoms of the issue include:

• Non-master signals constantly locking and unlocking

• Dropped frames in the affected sections of the display

• Tearing in the affected sections of the display

Possible root causes include the master signal or non-master signals not having stable syncs.
For more details, refer to the following:

• Signal Group (SST+SIGN?) section in the 

D4K Status System Technical Reference (P/N: 020-

101259-XX)

 and 

Mirage 4K Status System Technical Reference (P/N: 020-101450-XX)

.

• Troubleshooting > Incoming signal(s) problems section in the following manuals:

• D4K2560 User Manual (P/N: 020-101076-XX)

• Mirage 4K25 User Manual (P/N: 020-101361-XX)

• D4K3560 User Manual (P/N: 020-101075-XX)

• Mirage 4K35 User Manual (P/N: 020-101377-XX)

Solution

Change Slot 1 Input 1 cable (master) with the Slot 3 Input 1 cable. If the master signal is the issue, 
only Slot 3 Input 1 should show the lock/unlock symptom. If the video source is the issue, the 
symptoms should remain the same.

Screen has reversed colors

The screen has reversed colors (such as the blue and green are swapped) and the Formatter 
Mismatch alarm appears in the Status tab of the TPC. For example:

Details

The formatters are not connected correctly to the high speed image processor board (HIP)—the 
above example illustrates the green and blue formatters.
For more information, refer to the following:

• D4K Status System Technical Reference (P/N: 020-101259-XX)

• Mirage 4K Status System Technical Reference (P/N: 020-101450-XX)

• D4K2560 Interconnect Drawing (P/N: 020-101088-XX)

Formatter - Green HW Version

Mismatch Blue

Formatter - Blue HW Version

Mismatch Green

Содержание D4K Series

Страница 1: ...D4K and Mirage 4K Troubleshooting Te c h n i c a l R e f e r e n c e 020 101452 03...

Страница 2: ...g Problems or damage caused by improper installation alignment or by equipment modification if by other than Christie service personnel or a Christie authorized repair service provider h Problems or...

Страница 3: ...bol means that electrical and electronic equipment at their end of life should be disposed of separately from regular waste Please dispose of the product appropriately and according to local regulatio...

Страница 4: ...ayPort sources require reset when the projector is power cycled 7 Environmental Board HW Version status item displays Bad Power 8 Green vertical bars in lower right image 9 Dual 3G HD SD SDI Input car...

Страница 5: ...splayPort link connection which results in the graphics card driver dropping the signal Solution The latest AMD driver corrects this issue Contact your video source vendor for an update or download th...

Страница 6: ...ng Incoming signal s problems section in the following manuals D4K2560 User Manual P N 020 101076 XX Mirage 4K25 User Manual P N 020 101361 XX D4K3560 User Manual P N 020 101075 XX Mirage 4K35 User Ma...

Страница 7: ...rong temperature harness is being used for a particular formatter board For more information refer to the following D4K Status System Technical Reference P N 020 101259 XX Mirage 4K Status System Tech...

Страница 8: ...ironmental Board HW Version status item displays Bad Power At startup the Environmental Board HW Version status item displays Bad Power Details The environmental board EVB is not receiving 12V power f...

Страница 9: ...fore any boards are replaced If one or more of the signals are listed as No signal verify that the source is properly configured and the cables are secure between the source and the projector If one o...

Страница 10: ...aged 3G cards Ensure the mating BNC connector pin is not oversized Use 75 ohm connectors not 50 ohm Lamp hours not cleared A new projector does not have the lamp hours factory reset to zero Details St...

Страница 11: ...cting the TPC does not fix the Connection Lost message Power cycling the projector does not solve the problem The user cannot connect using the web page Serial commands using Ethernet RS232 RS422 do n...

Страница 12: ...for tilt angle which can accelerate the aging process of the lamp The projector line drawings were not available on the Christie website therefore the installers may not have been aware of the angle l...

Страница 13: ...oolant is flowing and no Temperature alarms have occurred the Flow Impeded alarm can be ignored as a temporary measure until the root cause for the error is found 4 Do not ignore other system generate...

Страница 14: ...D4K and Mirage 4K Troubleshooting D4K and Mirage 4K Troubleshooting Technical Reference 14 020 101452 03 Rev 1 10 2014 Solution Upgrade to the D4K60 Mirage 4K v1 3 0 software...

Страница 15: ...Seoul ph 82 2 702 1601 Japan Tokyo ph 81 3 3599 7481 India ph 91 080 6708 9999 Germany ph 49 2161 664540 France ph 33 0 1 41 21 44 04 Eastern Europe and Russian Federation ph 36 0 1 47 48 100 China Sh...

Отзывы: