5 Set-up and connection
Version 08/2009, Rev. 3.6 of 22/11/2021
• sb-bs
39 / 105
Translation of the original operating manual
Pos : 118 /200 Chris t/370 R VC-BA (ST ANDARDM ODU LE)/ 050 Aufst ellung und Anschl uss /050- 0060 Ansc hlus s v on Vak uum pumpe und/ oder Kühlf alle @ 25\m od_1404985088297_68. docx @ 187065 @ 2 @ 1
→
5.6
Connection of a vacuum pump and/or a cooling trap
In order to withdraw and condense the vapours that are formed, the RVC
can be connected with further components.
Pos : 119 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 120 /200 Chris t/371 R VC-BA (PR OJEKT E)/RVC 2- 33 CDpl us IR/ 050 Aufst ellung und Ansc hl uss/ 050- 0060- 0010 Abpum pen der Däm pf e über Vakuumpum pe RVC 2-33 IR @ 31\m od_1407484066961_68.docx @ 209056 @ 3 @ 1
→
5.6.1
Withdrawal of the vapours by a vacuum pump
The vapours are withdrawn by a vacuum pump, e.g. the vacuum diaphragm
pump for chemical applications of Vacuubrand, followed by condensation in
a liquid-cooled emission condenser. This application is suitable for low-
boiling samples containing solvents.
The vacuum pump must be connected to the RVC. The connector of the
stop valve must be plugged into the socket on the back of the unit (see
chapter 2.1.1 - "Functional and operating elements").
NOTE
The vacuum pump is supplied with power by the unit, but the maximum
current for the vacuum pump is limited. It is absolutely essential to refer to
the label of the electrical outlet for the vacuum pump (see the following
picture)!
If the current requirement of the vacuum pump is higher than the value
that is stated on the label, the pump must be supplied separately via an
on-site power socket.
1
Label indicating the
maximum current
Fig. 10: Indication of the maximum current for the vacuum pump (example)
The connector of the stop valve must be plugged into the socket "Stop
valve" on the back of the unit.
1
RVC
2
Vacuum sensor
3
Stop valve
4
Vacuum hose
5
Vacuum pump
Fig. 11: Combination of the RVC with a vacuum pump and a stop valve