Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
3 Safety
36 / 152
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
Translation of the original operating manual
Pos : 107 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/030 Sicherheit/ 030- 0020 Si cherheit Epsilon/ 030-0070- 0160 G ef ahr en durch Überdr uck ohne Ü berdrucksic her ung ( Standar d) @ 36\m od_1435915684375_68.docx @ 266991 @ 3 @ 1
3.7.16 Hazards caused by overpressure in the chamber
DANGER
Risk of injury caused by bursting of the chamber
The double chamber is designed for vacuum. It is not a pressure vessel!
In the event of a malfunction, an impermissible excess pressure may build
up in the chamber, which may cause the chamber to burst.
This may lead to life-threatening injuries.
• Do not use the freeze-dryer when it is pressurised (excess pressure)!
• We strongly recommend installing a pressure relief device (e.g. a
safety valve or rupture disc) with a sufficient rating.
Pos : 108 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 109 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/030 Sicherheit/ 030- 0020 Si cherheit Epsilon/ 030-0080 Sicherheits einric ht ung en Epsilon mit Netzsc halter, ohne T ürk ontak tsc halt er,Hy draulik @ 92 \mod_1657872575337_68. docx @ 962210 @ 2 @ 1
→
3.8
Safety devices
WARNING
The safety devices are aimed at protecting the operating personnel.
In case of safety devices that are not operating correctly or of safety-
relevant operating elements that are not freely accessible, there is a risk
of severe damage to health.
• Keep the safety-relevant operating elements freely accessible at all
times!
• Do not manipulate, remove, or disable the safety devices in any way!
• Check the correct operation of the safety devices at regular intervals in
accordance with the applicable national and international laws, rules,
and regulations concerning health and safety and the prevention of
accidents!
Control switch
Position see chapter 2.1.1 - "Functional and operating elements"
In order to disconnect the control circuit from the power supply, the control
switch must be switched off.
Equipotential bonding screw for earth conductor check
For the earth conductor check, there is an equipotential bonding screw on
the rear panel of the freeze-dryer. An earth conductor check can be carried
out with the aid of a suitable measuring instrument.
Pos : 110 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1