Chapter 3 Repair
111
Maintenance manual
MM-MIXO-EN Ind. 0
04/2012
Mixolab
■
Calibrating for 10 ml
1.
Go to the maintenance menu of the Mixolab.
2.
On the pump icon click on the corner coloured in brown (
A
).
3.
In the dialog box "Pump check/adjustment" enter the password
1576
, then validate by clicking on "OK".
4.
Click on Reset (
B
).
5.
Enter the password
2508
, then validate by clicking on "OK".
6.
The calibration parameters resetting window pops up. Click on
"Yes".
7.
Take a plastic vessel.
Put a piece of absorbing paper on the edge, attach it with a
piece of adhesive tape (
C
).
11 : Adjustments and checks for the Mixolab
Operation 11-9 Calibrating the Mixolab pump
Sheet: 1 / 4
Consumables:
Tools:
-
Precision scale in thousandths
B
A
C
Содержание Mixolab
Страница 2: ......
Страница 4: ......
Страница 6: ...Mixolab 6 Maintenance manual MM MIXO EN Ind 0 04 2012...
Страница 8: ...8 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 15: ...15 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 2 TROUBLESHOOTING...
Страница 16: ...16 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 31: ...31 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 3 REPAIR...
Страница 32: ...32 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 38: ...38 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 39: ...Chapter 3 Repair 39 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab 2 Instructions See following pages...
Страница 122: ...122 Chapter 3 Repair Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 123: ...123 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab Chapter 4 SPARE PARTS...
Страница 124: ...124 Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 126: ...126 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...
Страница 144: ...144 Chapter 4 Spare parts Maintenance manual MM MIXO EN Ind 0 04 2012 Mixolab...