Aqua-TR
72
User’s manual
07/2015
GB
■
Correcting the Hectoliter weight
In case of a systematic difference between the displayed Hectoliter
weight value and the reference Hectoliter weight, it is possible to
apply a corrective factor for this calibration (
HLW tolerance
±9,9 kg/hl).
After start-up or from the Test menu press key
, then use keys
and
to select the "
HLW" icon.
Press key
.
Select the calibration by mean of arrows
and
.
Press key
in order that the
HLW value is highlighted.
Use arrows
and
to modify the
HLW value.
Validate by pressing key
.
Press key
, to return to the main menu.