4-4
3. Output relay terminal connections
4. External drive connections and operation selection
The following operation can be executed by contact
signals from the terminals.
①
Selection of 3 chart speeds (speed numbers)
②
Recording execution and stop
③
Data printing
External drive functions and inter-terminal conditions
External drive functions
Across COM – EX
□
terminals
Remarks
ON : Shorted
OFF : Open
EX1
EX2
EX3
EX4
Chart speed 1
OFF
OFF
OFF
OFF
Receiving and feed operation by setting keys are effective.
For changing the selected chart speed number and chart
speed, refer to the instruction manual, option volume.
Chart speed 2
ON
OFF
OFF
OFF
Chart speed 3
OFF
ON
OFF
OFF
Recording execution
and stop
OFF
OFF
ON
OFF
Receiving and feed operation by setting keys are ineffective.
Data printing
OFF
OFF
OFF
ON
Keep ON time for longer than 1 second.
※)
Data cannot be printed unless recording is being executed.
Cautions
For the contacts being connected to the external drive terminals, use contacts of a switch, a relay, or the like
which is driven by lower than 30V AC or 60V DC or use manual contacts.
4-3-4 Alarm output terminals
(Caution 1) The alarm output is a mechanical relay “a” contact output. For connections, refer to
[3. Connections of output relay terminals].
Cautions
1. Connect a load of less than the designated contact
capacity to the output relay terminals.
2. Turn off the feed power source and buffer relay
power supply before starting connections for
preventing an electric shock accident.
①
Connect output relay terminals to the load via a
buffer relay.
②
For connection to the output terminals, use crimp
style terminals each having an insulation sleeve.
3. The buffer relay power supply is applied to the
output relay terminals after connections and an
electric shock accident occurs if you should touch
these terminals. Mount a terminal cover without fail
after connections.
b
a
Buffer relay
This unit
N.O
COM
:
Contact protective device
(Mounting on “a” side is desirable)
Power
supply
Load
Cautions
Turn off the feed power source before starting
connections of external drive terminals for
preventing an electric shock accident.
Terminal cover screw
(Remove two upper and
lower screws respectively)
メカリレー出力の接点容量
CH1
Power supply Resistive
load
Inductive
load
100VAC
0.5A
0.2A
240VAC
0.2A
0.1A
30VDC
0.3A
0.1A
CH12
Содержание LE5100