- 13 -
2. How to check
(1) Set a tester (standard DC voltage generator or precision variable resistor) to an input value equivalent to the
check target value.
(2) At this time, read the digital display and check if the error is within the specified accuracy range.
(3) Next, change to the check target channel and check it likewise.
(4) Check also the analog indication/dot printing position.
* The accuracy of the unit is rated at an ambient temperature of 23°C ±2°C. Ensure safety of the surrounding
environment.
* When you changed the wiring with the terminal cover removed, energize the unit for at least 30 minutes
after attaching the terminal cover and then perform checking.