- 12 -
5. Scale Adjustment
When you perform a trace printing (dot printing) position adjustment only, refer to “5-2. Zero and Span Adjustment for
Trace Printing”.
5-1. Scale
Check
The check is performed on a channel to channel basis. Even if all channels use the same measuring range, an error in
measured value may occur between channels.
The rated accuracy varies between digital display and
analog indication.
Use a tester with the accuracy equal to or higher than the
specified rated accuracy.
1. Preparation
(1) Turn OFF the power switch and perform the wiring depending on the input signal (see figure below).
Connect to the input terminal of the check target channel.
(2) Attach the terminal cover.
(3) Turn ON the power switch and change to [Continuous indication mode]. For details of [Continuous indication
mode], refer to “9. Continuous Indication Mode” in instruction manual “General”.
(4) Display the check target channel.
(5) Energize the unit for at least 30 minutes (an hour or more is ideal) and then perform checking.
Digital display: Temperature input ······· ±0.2%
DC voltage input ········ ±0.1%
Analog indication: ±0.5% of scale range
The accuracy of the unit is ±0.1%. Therefore, you
need to use a tester having higher degree of
accuracy to perform proper checking. Also, attention
should be paid to the thermocouple error.
Note that a tester requires time to ensure its stability
and accuracy.
Tester accuracy
Note 1
Make sure that the reference junction temperature is
0°C. When using an electronic reference junction
temperature compensator, read its instruction
manual. Also, the compensation accuracy should be
checked.
Reference junction compensator
Note 2