- 66 -
8-7. Chart Speed Settings “Chart”
Set the chart speed. When using remote contacts (optional), see also “13-1. External Operation Settings”.
(1) Pressing
the
key displays the
menu window (list of setting items).
(2) Select
“Chart”.
(3) 3-speed setting is available only when using
remote contacts (optional). Move the cursor
to the chart speed to be set, and press the
key to make it available for setting
and then enter a value.
Set a desired speed within the range of
(unit: 1mm) 1 to 600mm/H or 1 to 200mm/M,
but 12.5mm/H is an exception.
(4) After completing the settings of this item,
move the cursor to Set .
(5) Press
the
key to register the
settings (when chart recording is ON, a
setting change mark is printed). To cancel
the settings, press the
key.
[List of Chart setting parameters]
Parameter
Function
Default
Set value
ChartSpeed1
Set chart speed 1
25
1 to 600 (mm/H)
1 to 200 (mm/M)
ChartSpeed1unit
Select unit of chart speed 1
Mm/H
mm/H, mm/M
ChartSpeed2 (optional)
Set chart speed 2
25
1 to 600 (mm/H)
1 to 200 (mm/M)
ChartSpeed2 unit
Select unit of chart speed 2
mm/H
mm/H, mm/M
ChartSpeed3 (optional)
Set chart speed 3
25
1 to 600 (mm/H)
1 to 200 (mm/M)
ChartSpeed3 unit
Select unit of chart speed 3
Mm/H
mm/H, mm/M
Note: “ChartSpeed2” and “ChartSpeed3” are displayed
only when remote contacts (optional) are used.
All types of printing excluding time line printing, power-on printing, data printing, list printing and setting change mark will
be disabled. (See “6-3.3. Restrictions on recording”.)
Setting a speed at 150mm/H or higher
Note 1