Appendix 6 Mixer Measurement
391
Fig. 6.10 Mixer Mode Setting Dialog Box
Converter model
: set
[Input port]
,
[LO port]
and
[Output port]
of the mixer in the model dialog box. If the
LO and LO2 are not set, one external source can be used to provide the fixed LO; otherwise, the port must be
configured.
Fraction multiplier:
The (numerator/denominator) combination is a decimal and multiplied by the input and output
frequency. The default coefficient of the numerator/denominator of the multiplication base is 1, and the
adjustable range is 1 to 10. The negative is not allowed. If the multiplication coefficient changes, the
[Output
frequency range]
in the
[Frequency setting]
box will change accordingly. With the fraction multiplier, the
receiver frequency can be tuned to the harmonics of the mixer.
6.3 Calibration of Scalar Mixer Measurement
Perform calibration after mixer measurement setting. Menu path:
[Cal] > [Calibration...]
. Complete all
calibrations according to the guide.
1) Calibration setting
Fig. 6.11 Calibration Setting
a) Edit mixer
: click it to modify the mixer settings.
b) Calibrate the input and output end with the power meter (do not use the through type calibration mode).
If the through type calibration mode is unavailable, tick this option. In this case, the input and output end can be
calibrated with the power meter in the subsequent calibration.
c)Waveguide/In-fixer/On-wafer Setup
: the default mode is gray and unavailable.
2) Select the DUT connection type and calibration kit.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...