8 Basis of Network Measurement
8.3 Amplifier Parameter Specifications
249
The gain refers to the ratio between the output power (to the characteristic impedance load) and input power (from
the characteristic impedance source) of the amplifier and can be calculated by the following equation.
The output power of the amplifier is proportional to the input power in the case of small signal. The small-signal
gain is within the linear zone. With the power level of the input signal increasing, the amplifier will be saturated,
and the output power will reach the limit, leading to gain decline. The large signal gain is within the nonlinear zone.
See details in “
Appendix 5 Gain Compression Measurement of Amplifier
”.
8.3.2 Gain Flatness
The gain flatness refers to the gain change of the amplifier within the working frequency range. See details in “
8.10
Small Signal Gain and Flatness
”.
8.3.3 Reverse Isolation
Reverse isolation is applied to measure the transmission between the output end and input end. Similar to gain
measurement, only the stimulus signal is applied to the output end of the amplifier. See details in “
8.9 Reverse
Isolation
”.
8.3.4 Gain Drift Changes over Time (temperature and bias)
The gain drift refers to the maximum gain change over time, with the other parameters unchanged. It is a function
of the time. The gain drift related to other parameters can be measured, such as the gain deviation over the
temperature, humidity and bias voltage.
8.3.5 Linear Phase Deviation
Refer to the deviation of linear shifts. Ideally, the phase shift of the amplifier is a linear function of the frequency.
See details in “
8.8 Linear Phase Deviation
”.
8.3.6 Group Delay
The group delay refers to the transition time of the signal through the amplifier. It is a function of the frequency.
The ideal linear phase shift has a constant change rate over the frequency. In this case, the group delay is a constant.
In the actual measurement, the group delay is calculated by the following equation:
See details in “
8.5 Group Delay
”.
8.3.7 Return Loss (standing wave ratio, ρ)
The return loss refers to the reflection matching of the amplifier relative to the system impedance Z0 in
measurement of the input or output port. It can be calculated by the following equation:
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...