5.6
Troubleshooting
No.
Trouble phenomenon
Cause
Remedy
①
Organic EL display
shows nothing.
Power not turned on
Turn on the power
Blown fuse
Replace the fuses.
②
Motor fails to start or
suddenly stops
running.
Faulty motor
Call for repair.
The motor must be replaced.
Stopped due to
overload or abnormal
temperature
[1] Check that the exhaust/suction ports are
not blocked.
[2] Check the rated voltage.
[3] Check for multiple (octopus) connections.
[4] Check for motor overheat due to either
filter clogging or suction temperature.
After the checks [1] to [4], take corrective
action, and then turn the main power switch
off and back on.
If the operation cannot resume, the
temperature thermostat of the motor may be
active. Turn off the main power, and after 30
minutes, resume the operation.
Primary or secondary
filter not in place
Install the primary and secondary filters
correctly.
Snap locks
disengaged
Engage all the snap locks firmly.
③
Lower suction force
Clogged filter
Replace filters.
If filters are left clogged, the device may result
in “untrapped particles” or “temperature rise in
dust collector”.
Clogged piping or
suction port
Check if piping is clogged or the suction port
blocked.
Faulty motor
Call for repair.
The motor must be replaced.
④
Untrapped particles
Improper filter
installation
Reinstall filters.
Filter broken or past
service life
Replace filters.
Clogged filter
Replace filters.
⑤
Odd noise or
vibration from motor
Foreign matter
entered in blower
Call for repair.
Broken motor bearing Call for repair.
Note: For other phenomena, contact CHIKO AIRTEC.
Содержание CBA-750AT3-HC-DSA-V1
Страница 5: ...1 CMN034 002 Translation of the original instructions 3 1 1 1 1 2 10 60 80 1 3 0 40 80 100 1 000...
Страница 6: ...1 4 1 4 2 OFF 1 5...
Страница 7: ...1 CMN034 002 Translation of the original instructions 5 1 6...
Страница 8: ...2 6 2 2 1 100V 200V T 220 230V CE 1 1 2 1 1 HEPA 1 2 3m 100V 1 200V T 1 220 230V CE 1 1 100 115V...
Страница 10: ...2 8 2 3 AT3 EL OLED No 1 7 2 4 9 4 14 1 1 1 1 OFF 3 3 3 13 ON ENTER 3 3 3 13 4 14 MODE SELECT 4 14...
Страница 12: ...2 10 2 4 3 OP kPa SP kPa DP kPa EP kPa Blower Runtime h 17 Total h Moter...
Страница 13: ...3 CMN034 002 Translation of the original instructions 11 3 3 1 3 1 1 0 40 80 RH REMOVE 1 1 1...
Страница 14: ...3 12 3 1 2 10 1 2 3 2 1 ON RS485 ID 2 AT3 ON 3 4 1 7 3 ON OFF 30 3 ON OFF 1 1 Ver ID...
Страница 20: ...5 18 5 5 1 WARN2 1 5 1 1 1 1 1 2 2 1 3 1 4 2 1...
Страница 22: ...5 20 3 4 5 HEPA 6 5 2 1 Littell fuse 218 250V 10A 1 2 5 3 1 1 1 1 5 4...
Страница 26: ...6 24 6 6 1 6 1 1...
Страница 27: ...6 CMN034 002 Translation of the original instructions 25 1 5V 0 2V DC 1 2...
Страница 31: ...7 CMN034 002 Translation of the original instructions 29 7 3 7 3 1 CBA 750AT3 HC DSA V1...
Страница 32: ...30 12 7 2 7 2 28...
Страница 63: ...MEMO...