1
Operating Instructions
Tachometer Stroboscope Calibrator / Verifier
TSV-1000
Страница 1: ...1 Operating Instructions Tachometer Stroboscope Calibrator Verifier TSV 1000 ...
Страница 2: ... keep it nearby for future reference SPECIFICATIONS Accuracy 0 01 1 digit Presets 10 Programmable RPM setpoints TACHOMETER MODE Operating Modes TACHOMETER and STROBOSCOPE Power Supply 100 240V AC to 5V DC Warranty 1 Year BEFORE USING THE TSV 1000 Plug in the AC 100 240VAC to DC 5V power adapter Remove the pulsor sensor from the rear panel stor age slot and plug it into the front panel USB connect ...
Страница 3: ...eter is in measuring mode its display should be in sync or close to in sync with TSV 1000 s display The allowable tolerance will depend on the ta chometer s specifications While in TACHOMETER mode press SP or SP to toggle between presets 10 Press ADJ or ADJ to adjust the pulse rate To adjust the presets press DEFAULT the display will show Def X Y where X is the preset number 0 9 and Y is the prese...
Страница 4: ...ester 3 24 inches 75 600mm away from the TSV 1000 pulse sensor The TSV 1000 s display should be in sync or close to stroboscope s display The allowable tolerance will depend on the stroboscope s specifica tions In STROBOSCOPE mode the default display is FPM flashes per minute To change the display to HZ press and hold S for 3 seconds It will now display HZ To re turn to FPM press and hold S for 3 ...