1. PRESENTATION
1.1. GENERAL
MiniFlex
®
sensors take the form of a flexible coil connected by a shielded lead to a housing containing
the signal processing electronics, powered by batteries.
The flexibility of the sensors facilitates the clamping of the conductor to be measured, whatever its nature
(cable, bar, strand, etc.) and its accessibility. The design of the coil opening and closing device, by snap
locking, allows handling while wearing protective gloves.
The electronic unit can be connected to the mV
ac
or V
ac
input of a measuring instrument having an input
impedance of ≥ 1 M
W
.
1.2. SINGLE-PHASE MINIFLEX
®
MiniFlex
®
series MA110 single‑phase sensors constitute a line of three models, 17, 25, and 35 cm long,
that convert alternating currents ranging from 3 to 3000 A into proportional AC voltages.
OFF
3 A
30 A
300 A
3000 A
3 A : 1000 mV / A
30 A : 100 mV / A
300 A : 10 mV / A
3000 A : 1 mV / A
AC CURRENT PROBE
MA 110
600 V CAT IV
1000 V CAT III
On
Auto Power Off
OL
P
Flexible coil.
Sensor opening device.
Shielded cord.
2 banana plugs, dia. 4 mm.
Switch to activate the
instrument and choose
the sensitivity.
Type B Micro-USB port
to connect a 5V power
supply.
Connecting cord between the
electronic unit and a display
device.
Red OL indicator reporting an
overshoot of the measurement
range (overload).
Green On indicator indicating
that the device is on (blinking)
or connected to the optional
mains adapter (steady).
Orange Auto Power off
indicator indicating that
automatic standby switching
of the device is activated.
Auto Power Off
button
to deactivate automatic
switching off.