30
Read the measurement on the multimeter, applying the reading coefficient indicated on the label on the
electronic unit corresponding to the setting of the switch.
3 A~ range
1000 mV~/A~
30 A~ range
100 mV~/A~
300 A~ range
10 mV~/A~
3000 A~ range
1 mV~/A~
Multiply the reading by the coefficient.
For example, a reading of 1 V on the measuring instrument corresponds to a current of = 100 A
in the 300 A~ range.
2.2.2. DISCONNECTING THE MA130
Withdraw the flexible coil from the conductor, set the switch to
OFF,
and disconnect the electronic unit
from the multimeter.
2.2.3. IDENTIFYING THE SENSORS OF THE MA130
To identify the sensors and the output leads, you can mark them with the coloured rings provided with
the device.
Clip rings of the same colour on the sensor and on the corresponding output leads.