4.2.1.
HOLD key: Management and hold of the display
Three operating modes are possible:
-
the
RUN
mode
HOLD inactive
-
the
HOLD
mode
[F2]
-
the
AUTO HOLD
mode
[F3]
The
HOLD
mode freezes the screen with the main measurement in progress
and the history curve as they were when the HOLD key was pressed. The instrument
continues to manage the measurements on the secondary display unit (
REL
mode).
The type of range selection remains the same: AUTO or MANUAL depending
on the configuration when this mode was entered.
The curve resumes at zero when RUN is pressed.
The AUTO HOLD mode automatically freezes on the screen the main
measurement in progress each time a stable measurement is detected. It is confirmed
by the emission of an audible beep (if the "No beep" configuration was not selected in
the Configuration menu).
The stored values remain displayed until the next stable measurement is made
(measurement different by ±100 digits) or until the
AUTO HOLD
mode is exited by
RUN
.
The instrument continues to manage the measurements and to display them in the
graphic window or on the secondary display unit (
REL
mode).
The type of range selection remains the same (in AUTO or MANUAL)
depending on the configuration when this mode was entered. The AUTO HOLD mode
is available only for V and A measurements.
4.2.2.
MEAS key: Advanced measurements