18
Typical noise level on output (peak‑to‑peak value)
Frequency band
DC - 100kHz
10mV/A range
480µV
100mV/A range
3mV
These data assume an ambient temperature of 23±3°C, a relative humidity of 20 to 75%, a frequency
from DC to 1kHz, a load impedance of 1M
W
/100pF, and a conductor that is centred and parallel to
the mark.
Pass band: DC to 100kHz.
The pass band of the associated oscilloscope depends on the frequency of the signal to be measured.
A pass band greater than four times the frequency of the signal to be measured is sufficient.
Frequency of use (not entailing an additional error of more than 3% with respect to the reference
Intrinsic uncertainty (%)
10
40
20
50
30
60
-15
-10
-5
0
0
80
90
70
100
I (A)
-20
4. SPECIFICATIONS
4.1. ELECTRICAL CHARACTERISTICS
Range
Measurement range
Intrinsic uncertainty
100mV/A range
50mA to 10A peak
3% +50mA
10mV/A range
50mA to 40A peak
4% +50mA
10mV/A range
40A to 100A peak
See curve below
Linearity for a DC signal (10mV/A range)