50
Conformity
9.4.
9.4.1.
Safety
According to NF EN 61010-1:
Insulation
class 2
Degree of pollution
2
Use indoor
Altitude
< 2000 m
Measurement categoryof the "measurement" inputs CAT III, 1000V with respect to earth
Measurement categoryof the "measurement" inputs CAT IV, 600V with respect to earth
9.4.2.
EMC
This instrument is designed in conformity with the EMC standards in force and its compatibility has been tested in
accordance with the following standards:
Emissions (cl. A) and Immunity
NF EN 61326-1
9.4.3.
Radio Equipment Directive (RED)
– for the BT versions of the multimeters
This instrument is designed in compliance with directive 2014/53/EU and has been tested in accordance with the following
standards:
ETSI EN 301 489-1
ETSI EN 301 489-17
ETSI EN 300 328
EN 62311
10. MECHANICAL SPECIFICATIONS
Housing
10.1.
Dimensions
196 x 90 x 47.1mm
Mass
570g
Materials
ABS V0
Dust- and water-tightness IP67,
according to NF EN 60529 (Not in operation; if the instrument is
immersed, it must be dried, in particular the terminal block, before it can be used again)
11. MAINTENANCE
Except for the fuse and the batteries, the instrument contains no parts that can be replaced by personnel who have
not been specially trained and accredited. Any unauthorized repair or replacement of a part by an "equivalent" may
gravely impair safety.
Cleaning
11.1.
Disconnect the instrument completely and switch it OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or forced air. Do not
use alcohol, solvents, or hydrocarbons.
Make sure that no foreign body interferes with the operation of the snap device of the sensor.
Replacement of the fuse
11.2.
For safety reasons the fuse must always be replaced by an identical model:
11A: 10x38 -1,000V -F
Updating of the internal software
11.3.
With a view to providing, at all times, the best possible service in terms of performance and technical upgrades, Chauvin
Arnoux invites you to update the embedded software of the device by downloading the new version, available free of charge
on our web site.
Our site:
http://www.chauvin-arnoux.com Embedded software
Under
Support
, click on
ASYC IV
e
mbedded software /Loader Asyc IV v.xx.exe