34
5.9.2.
Main measurement
To measure a temperature:
1.
Connect the sensor to the V and COM terminals, making sure to get the polarity right.
2.
Choose the unit: °C (Celsius), K (Kelvin) or °F (Fahrenheit).
3.
Select ".../…".
4.
Choose the type of sensor Pt100, Pt1000, TCJ, or TCK.
If "
OL
" is displayed, the sensor is open-circuit or the measured value exceeds the range limit.
2 presses T°
Measurement of the temperature with a sensor: Pt100/Pt1000.
“Active” protection by PTC thermistor To connect a 2-wire PT probe to the multimeter, we
recommend the use of the PT100 probe module
HX0091.
3 presses T°
Measurement of the temperature in °Celsius using a thermocouple between the V and COM
terminals
K thermocouple from -40°C to +1200°C or TCJ (J thermocouple) from -40°C to +750°C
Without a TK thermocouple, you can determine the ambient temperature inside the multimeter
with a bridge between the V and COM terminals.
The
keys of the navigator are used to change the scale of the graphic window.
The scale selected is transcribed in the help line.
Measurement on an MLI type speed variator
5.10.
5.10.1.
Connecting the multimeter
Connecting the multimeter to filter a voltage >300 Hz