3.3.3. DIODE
Particular reference conditions:
Zero voltage.
Diode without resistance or capacitance in parallel.
Diode voltage measured between 0.29 and 2V.
3.3.4. CAPACITANCE
Particular reference conditions:
Zero voltage.
Capacitance without resistance in parallel.
Range
3 nF *
30 nF *
300 nF
3 µF
Measurement
range
400 pF to
2.999 nF
3.00 to
29.99 nF
30.0 to
299.9 nF
0.300 to
2.999 µF
Resolution
0,001 nF
0,01 nF
0,1 nF
0,001 µF
Intrinsic uncer-
tainty
5% ± 10 pt
5% ± 5 pt
*: In these ranges, subtract the no-load value (typically 50pF)
from all readings.
Range
30 µF
300 µF
3 mF
30 mF
Measurement
range
3.00 to
29.99 µF
30.00 to
299.9 µF
0.300 to
2.999 mF
3.00 to
29.99 mF
Resolution
0,01 µF
0,1 µF
0,001 mF
0,01 mF
Intrinsic uncer-
tainty
5% ± 5 pt
The 3mF and 30mF ranges are not available in automatic
mode.
3.3.5. NON-CONTACT VOLTAGE DETECTION (NCV)
The instrument detects the line voltage at 230VAC with respect
to ground, at 50 Hz and at a distance of less than 5cm.
3.3. ENVIRONMENTAL CONDITIONS
Operating range:
-10°C at 55°C and
≤
80%RH without condensation up to 40° C.
Storage range (without battery):
-20°C at +55°C and
≤
90%RH without condensation up to 45°C.
If an extended period of non-use is anticipated, or for storage,
withdraw the batteries from the housing.
For use indoors and outdoors without rain.
Pollution degree: 2.
Altitude: <2000m.
3.4. POWER SUPPLY
The instrument is powered by two 1.5V alkaline batteries
(type AAA or LR3).
Battery life is 100 h.