3. RESISTANCE MEASUREMENT
1) Connect the 2 cables to the 4 measurement terminals, then the 2 Kelvin clips to the object to be
tested, which must not be live.
2
C
2
P
1
P
1
C
2) Set the switch to 400
W
- 10 mA.
A
Ω
+I
START/STOP
3) Start the measurement
by pressing the START/
STOP button.
If the measurement is too low, turn
the switch to the next lower range
and restart the measurement.
Continue until the display shows
at least 3 digits.