Chattanooga Intelect OPERATING INSTRUCTIONS 

 
 

Ultrasound 

1.

 

Turn on unit (the main switch is on top of the main 
screen on the left) 

2.

 

Select “Ultrasound” menu (this is the 2

nd

 key on the 

console (first row)).  

3.

 

Press “Customize” 

4.

 

Set Frequency: 1 or 3 MHz 

5.

 

Set Mode: Continuous or Pulse mode (duty cycle 
10%, 20%, or 50%) 

6.

 

Set display (W or W/cm

2

). Ensure that it is selected 

to display W/cm

2

 

7.

 

Set time 

8.

 

Ignore Duty Cycle Frequency 

9.

 

Place the play button on the round dial.  

10.

 

Apply gel to the tissue and then apply the 
transducer. To start the treatment, rotate the 
Intensity knob to the desired output. The light on the 
ultrasound head will turn green when the unit is 
active.  

11.

 

If you want to stop treatment before treatment is 
over, turn the dial down to 0, or press the stop 
button below the round dial. You can also pause 
treatment by pressing the pause button on the 
round dial. 

12.

 

When treatment is completed, hold down the power 
button until the unit turns off  

 

TENS 

1.

 

Turn on unit (the main switch is on top of the 
main screen on the left) 

2.

 

Press Channel 1. 

3.

 

Select “Electrotherapy”. This is the on the 1

st

 row 

of buttons on the console 

4.

 

Select “Pulsed Biphasic”  

5.

 

Select “symmetrical biphasic TENS” 

6.

 

Press customize: 

a.

 

Adjust the phase duration  

b.

 

Ignore frequency modulation  

c.

 

Ignore amplitude modulation 

d.

 

Adjust the frequency  

e.

 

Ignore beat frequency 

f.

 

Adjust time 

g.

 

Leave to CC 

h.

 

Adjust the amplitude 

i.

 

Once desired amplitude is reached, press 
the play button 

7.

 

Press the back button 

8.

 

Press channel 2 

9.

 

Repeat steps 3-6 

10.

 

Press channel 1 

11.

 

Press play 

12.

 

When treatment is completed, hold down the 
power button until the unit turns off 

 

 
 

 