15 700 02 0319
Device Description
13
•
Insert the projectile into the fitted guide tube.
•
Screw the shaft onto the handpiece until fin
-
ger-tight.
• Tighten the shaft using the supplied open-end
spanner. It must no longer be possible to unscrew
the shaft by hand.
• Make sure that the cap parts of the pulse transmitter
are screwed firmly in place and that the pulse trans
-
mitter screw cap is screwed firmly to the shaft.
• Insert the pulse transmitter insert into the corre-
sponding pulse transmitter screw cap.
• Screw the pulse transmitter screw cap onto the
handpiece until finger-tight.
NOTE
Depending on the pulse transmitter, the pulse transmitter screw cap is in one
or two pieces.
Here, the images show how to assemble the handpiece with a two-piece pulse
transmitter screw cap.
3.5.3
Connecting the electrical power supply
• Connect the mains cable to the mains connector on the left rear of the device
(see Fig. 3-2).
• Insert the mains cable into the electrical socket.
ATTENTION!
Maintain a minimum distance between the device and the wall so that the
mains plug can be pulled out without restriction (disconnected from the
power supply network) and the ventilation slits on the rear are not blocked.
The device must only be connected to properly earthed and correctly installed
shockproof sockets!