2 SAFETY PRECAUTIONS
2.2 PRECAUTIONARY INSTRUCTIONS (continued)
C. Dangers
DO NOT connect the unit to an electrical
supply without first verifying that the
power supply is the correct voltage.
Incorrect voltage may cause unit damage,
malfunction, electrical shock, fire, or
personal injury. Your unit was constructed
to operate only on the electrical voltage
specified on the Voltage Rating and Serial
Number Plate. Contact the Chattanooga
Group dealer if the unit is not properly
rated.
Power Supplies retain High Voltage!
Discharge the C4 Capacitor prior to
complete removal of the Power Supply.
Discharge the C4 Capacitor prior to
handling of the new Power Supply.
Triton® is a prescription device to be used
only under the supervision of and by the
order of a physician or other licensed health
care provider.
Do not use any other knot styles in the
Cord other than those specified in this
manual. Use of other knot styles could
cause Cord to completely disengage from
the system and may cause serious injury
to patient and personnel.
•
•
•
•
•
•
Use only accessories that are specially designed
for this unit. Do not use accessories manufactured
by other companies on this (table, unit, device,
etc.). Chattanooga Group is not responsible for
any consequence resulting from using products
manufactured by other companies. The use
of other accessories or cables may result in
increased emissions or decreased immunity of
this (table, unit, device, etc.).
This component removal and replacement
procedure requires that the following be performed
after component or assembly replacement to
ensure proper alignment and operation of the unit:
COMPLETE CALIBRATION OF UNIT
Refer to Section 7
AND
MINIMUM OF 2 HOUR BURN IN OF UNIT AFTER
SUCCESSFUL CALIBRATION Refer to Section 7
Failure to perform either of these procedures, after
a component or assembly replacement will void
warranty and may result in injury to patients
and/or severe equipment damage.
It is not recommend tying a new knot in a cord that
has frayed or failed in use or is worn in an area from
constant use. The cord must be replaced.
•
•
•
Содержание TRITON 2841
Страница 15: ...13 Triton Traction Unit NOTES ...
Страница 72: ...Triton Traction Unit 70 9 SCHEMATICS Control Board 47801 1 of 3 ...
Страница 73: ...Triton Traction Unit 71 9 SCHEMATICS Control Board 47801 2 of 3 ...
Страница 74: ...Triton Traction Unit 72 9 SCHEMATICS Control Board 47801 3 of 3 ...
Страница 75: ...Triton Traction Unit 73 9 SCHEMATICS Motor Control Board 47803 1 of 2 ...
Страница 76: ...Triton Traction Unit 74 9 SCHEMATICS Motor Control Board 47803 2 of 2 ...
Страница 77: ...Triton Traction Unit 75 9 SCHEMATICS Patient Interrupt Switch 48056 1 of 1 ...
Страница 78: ...Triton Traction Unit 76 9 SCHEMATICS Serial Interface 47022 1 of 1 ...
Страница 79: ...Triton Traction Unit 77 9 SCHEMATICS Smart Card 47807 1 of 1 ...
Страница 80: ...Triton Traction Unit 78 9 SCHEMATICS sEMG Module 48123 1 of 1 ...
Страница 81: ...Triton Traction Unit 79 9 SCHEMATICS Block Diagram 1 of 1 ...
Страница 82: ...Triton Traction Unit 80 10 SPECIAL FIXTURES Calibration Fixture 1 of 4 ...
Страница 83: ...Triton Traction Unit 81 10 SPECIAL FIXTURES Calibration Fixture 2 of 4 ...
Страница 84: ...Triton Traction Unit 82 10 SPECIAL FIXTURES Calibration Fixture 3 of 4 ...
Страница 85: ...Triton Traction Unit 83 10 SPECIAL FIXTURES Calibration Fixture 4 of 4 ...