11
Triton® Traction Unit
ERROR
CODE
ERROR
TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
ERROR MESSAGES (200-223) REQUIRING TECHNICAL ASSISTANCE (continued)
214
ERROR
User has confi gured a treatment such that it takes
too much memory to compute the treatment
steps. Unit will default back to the default power
up protocol.
Under normal operating conditions, this Error should
never appear.
Remove unit from service and report exact treatment
parameters to Chattanooga Group.
215
ERROR
Internal error, unit cannot get the next treatment
step information, treatment is terminated.
Multimedia Card (MMC) Data is corrupt.
Multimedia Card (MMC) interface not working.
1.
Try known good Multimedia Card (MMC).
2.
Remove unit from service and report exact treatment
parameters to Chattanooga Group.
217
ERROR
Error reading a MMC.
Bad Mutimedia Card (MMC).
Improper cable connection to Motor Control Board.
Bad Touch Screen Assembly.
1.
Try known good Multimedia Card (MMC).
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Touch Screen Assembly.
218
ERROR
Error upgrading Control Board software.
Bad Mutimedia Card (MMC).
Improper cable connection to Motor Control Board.
Bad Touch Screen Assembly.
1.
Try known good Multimedia Card (MMC).
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Touch Screen Assembly.
219
ERROR
Error upgrading Control Board software.
Bad Mutimedia Card (MMC).
Improper cable connection to Motor Control Board.
Bad Touch Screen Assembly.
1.
Try known good Multimedia Card (MMC).
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Touch Screen Assembly.
220
ERROR
Error upgrading Motor Control Board software.
Bad Mutimedia Card (MMC).
Improper cable connection to Motor Control Board.
Bad Motor Control Board.
Bad Touch Screen Assembly.
1.
Try known good Multimedia Card (MMC).
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Motor Control Board.
4.
Replace Touch Screen Assembly.
221
ERROR
Error upgrading Motor Control Board software.
Bad Mutimedia Card (MMC).
Improper cable connection to Motor Control Board.
Bad Motor Control Board.
Bad Touch Screen Assembly.
1.
Try known good Multimedia Card (MMC).
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Motor Control Board.
4.
Replace Touch Screen Assembly.
222
ERROR
Patient switch is shorted.
Patient Interrupt Switch not completely seated in
connector.
Bad Patient Interrupt Switch.
Bad Patient Interrupt Switch Connector.
1.
Properly connect Patient Interrupt Switch to unit.
2.
Use a known good Patient Interrupt Switch.
3.
Replace Patient Interrupt Switch Connector on unit.
223
ERROR
Beta version evaluation period has expired.
Unit is designated BETA Test Unit.
Immediately return unit to Chattanooga Group.
5.1 TRITON ERROR MESSAGES (continued)
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga
Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field
Service Technician certified by Chattanooga Group before any further operation or use of the unit. Use of a unit that indicates an Error or Warning in these
categories may pose a risk of injury to the patient, user, or may cause extensive internal damage to the unit.
5 TROUBLESHOOTING
Содержание TRITON 2841
Страница 15: ...13 Triton Traction Unit NOTES ...
Страница 72: ...Triton Traction Unit 70 9 SCHEMATICS Control Board 47801 1 of 3 ...
Страница 73: ...Triton Traction Unit 71 9 SCHEMATICS Control Board 47801 2 of 3 ...
Страница 74: ...Triton Traction Unit 72 9 SCHEMATICS Control Board 47801 3 of 3 ...
Страница 75: ...Triton Traction Unit 73 9 SCHEMATICS Motor Control Board 47803 1 of 2 ...
Страница 76: ...Triton Traction Unit 74 9 SCHEMATICS Motor Control Board 47803 2 of 2 ...
Страница 77: ...Triton Traction Unit 75 9 SCHEMATICS Patient Interrupt Switch 48056 1 of 1 ...
Страница 78: ...Triton Traction Unit 76 9 SCHEMATICS Serial Interface 47022 1 of 1 ...
Страница 79: ...Triton Traction Unit 77 9 SCHEMATICS Smart Card 47807 1 of 1 ...
Страница 80: ...Triton Traction Unit 78 9 SCHEMATICS sEMG Module 48123 1 of 1 ...
Страница 81: ...Triton Traction Unit 79 9 SCHEMATICS Block Diagram 1 of 1 ...
Страница 82: ...Triton Traction Unit 80 10 SPECIAL FIXTURES Calibration Fixture 1 of 4 ...
Страница 83: ...Triton Traction Unit 81 10 SPECIAL FIXTURES Calibration Fixture 2 of 4 ...
Страница 84: ...Triton Traction Unit 82 10 SPECIAL FIXTURES Calibration Fixture 3 of 4 ...
Страница 85: ...Triton Traction Unit 83 10 SPECIAL FIXTURES Calibration Fixture 4 of 4 ...