10
Triton® Traction Unit
5 TROUBLESHOOTING
ERROR
CODE
ERROR
TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
ERROR MESSAGES (200-223) REQUIRING TECHNICAL ASSISTANCE
200
ERROR
The motor PCB has reset.
Motor Control Board is bad.
If the Error repeats routinely, replace Motor Control Board.
201
ERROR
Error communicating with the realtime clock
chip.
Control Board Communication Error.
If the Error repeats routinely, replace Touch Screen Assembly.
202
ERROR
The motor PCB returned an error while
performing the zero Load Cell function.
Bad Load Cell Assembly
Motor Control Board Communication Error.
1.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
2.
Replace Load Cell Assembly.
3.
Replace Motor Control Board.
4.
Replace Touch Screen Assembly.
203
ERROR
Timeout waiting for the motor PCB to perform
the zero Load Cell calibration.
Bad Load Cell
Bad Motor Control Board
1.
Ensure there is slack in the Cord.
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Load Cell.
4.
Replace Motor Control Board.
204
ERROR
Timeout waiting for the motor PCB to perform
the motor current calibration.
Motor Control Board Communication Error.
1.
Ensure there is no slack in the Cord.
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Load Cell.
4.
Replace Motor Control Board.
205
ERROR
A Motor fault error has occurred while a
treatment was running.
Internal mechanical jam of unit.
Bad Motor Control Board or Motor Assembly.
1.
Correct any internal mechanical jam.
2.
Replace bad Motor Control Board or bad Motor Assembly.
206
ERROR
The Motor PCB cannot get to the desired tension.
Internal Mechanical Jam of unit.
Bad Motor Control Board or Motor Assembly.
1.
Correct any internal mechanical jam.
2.
Replace bad Motor Control Board or bad Motor Assembly.
207
ERROR
We have detected a motor current error
(MC_MON pin on motor PCB).
Unit out of calibration.
Perform Unit Calibration and Burn In procedures.
208
ERROR
We have detected a Load Cell limit error
(LC_MON pin on motor PCB).
Unit is out of calibration.
Bad Motor Control Board.
Bad Load Cell Assembly.
1.
Perform Unit Calibration and Burn In procedures.
2.
Replace Motor Control Board.
3.
Replace Load Cell Assembly.
209
ERROR
Error erasing a patient card.
Bad Patient Data Card.
Bad Touch Screen Assembly.
1.
Try known good Patient Data Card.
2.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
3.
Replace Touch Screen Assembly.
210
ERROR
CRC not correct on treatment just read from
Patient Card.
Bad Patient Data Card.
Bad Touch Screen Assembly.
1.
Try known good Patient Data Card.
2.
Replace Touch Screen Assembly.
211
ERROR
Error reading default protocol on power up so
must rebuild protocols to factory presets.
EEPROM is corrupt.
If the Error repeats routinely, replace Touch Screen Assembly.
212
ERROR
The Motor PCB fl agged a comm watchdog error.
Touch Screen Assembly did not poll Motor Control
Board in a reasonable amount of time.
1.
Check Cables and Harnesses for proper seating and
connection on Motor Control Board.
2.
If the Error repeats routinely, replace Touch Screen
Assembly.
213
ERROR
Motor PCB has indicated that the zero calibration
value is invalid and needs to be calibrated.
Unit is out of calibration.
Perform Unit Calibration and Burn In procedures.
5.1 TRITON ERROR MESSAGES (continued)
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga
Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field
Service Technician certified by Chattanooga Group before any further operation or use of the unit. Use of a unit that indicates an Error or Warning in these
categories may pose a risk of injury to the patient, user, or may cause extensive internal damage to the unit.
Содержание TRITON 2841
Страница 15: ...13 Triton Traction Unit NOTES ...
Страница 72: ...Triton Traction Unit 70 9 SCHEMATICS Control Board 47801 1 of 3 ...
Страница 73: ...Triton Traction Unit 71 9 SCHEMATICS Control Board 47801 2 of 3 ...
Страница 74: ...Triton Traction Unit 72 9 SCHEMATICS Control Board 47801 3 of 3 ...
Страница 75: ...Triton Traction Unit 73 9 SCHEMATICS Motor Control Board 47803 1 of 2 ...
Страница 76: ...Triton Traction Unit 74 9 SCHEMATICS Motor Control Board 47803 2 of 2 ...
Страница 77: ...Triton Traction Unit 75 9 SCHEMATICS Patient Interrupt Switch 48056 1 of 1 ...
Страница 78: ...Triton Traction Unit 76 9 SCHEMATICS Serial Interface 47022 1 of 1 ...
Страница 79: ...Triton Traction Unit 77 9 SCHEMATICS Smart Card 47807 1 of 1 ...
Страница 80: ...Triton Traction Unit 78 9 SCHEMATICS sEMG Module 48123 1 of 1 ...
Страница 81: ...Triton Traction Unit 79 9 SCHEMATICS Block Diagram 1 of 1 ...
Страница 82: ...Triton Traction Unit 80 10 SPECIAL FIXTURES Calibration Fixture 1 of 4 ...
Страница 83: ...Triton Traction Unit 81 10 SPECIAL FIXTURES Calibration Fixture 2 of 4 ...
Страница 84: ...Triton Traction Unit 82 10 SPECIAL FIXTURES Calibration Fixture 3 of 4 ...
Страница 85: ...Triton Traction Unit 83 10 SPECIAL FIXTURES Calibration Fixture 4 of 4 ...