9
Triton® Traction Unit
5.1 TRITON ERROR MESSAGES
5 TROUBLESHOOTING
A.
The following information is provided
as an aid in defining the Software Error
Messages of the Triton traction unit.
Once a particular Error Message is defined,
the information will also list probable causes
and possible remedies. Once the problem
area is determined, subsequent tests for
verification will be necessary to determine
the subassembly or PC Board to be replaced.
In the case of PC Boards, all troubleshooting
and tests will be to validate a “Bad Board”
only. No component level troubleshooting
information is or will be provided by
Chattanooga Group for field troubleshooting
of PC board components.
B.
Once a particular subassembly or PC Board
has been determined as bad, refer to the
appropriate Removal & Replacement
Section for the board affected and follow the
instructions for replacement of the board.
ERROR
CODE
ERROR
TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
USER CORRECTABLE WARNING MESSAGES
100
WARNING
Treatment has been running for 8 seconds,
but no tension is detected on the Cord.
Too much slack in Traction Cord
Remove Traction Cord Slack.
101
WARNING
Patient pressed Patient Switch.
Patient Interrupt Switch button pressed.
Touch the Touch Screen to clear message.
102
WARNING
Patient Switch is unplugged.
Patient Interrupt Switch not properly connected to unit.
Properly connect Patient Interrupt Switch and touch Touch
Screen to clear message.
104
WARNING
User selected Patient Card button on
utilities screen, but no traction treatments
were found on the card.
Wrong or bad Patient Data Card inserted into unit.
Insert correct Patient Data Card.
105
WARNING
User selected save to card, but no card is
inserted into the unit.
No Patient Data Card inserted.
Properly insert correct Patient Data Card into unit.
106
WARNING
User selected save to card, but the card
currently inserted is not a Patient Card.
Wrong or bad Patient Data Card inserted into unit.
Insert correct Patient Data Card.
107
WARNING
User selected save to card, but the card
currently inserted is full.
Memory full on Patient Data Card used.
Save Data to PDMS and erase Patient Data Card.
108
WARNING
User typed in a blank patient name.
No Patient ID assigned.
Enter Patient ID.
109
WARNING
User typed in a blank protocol name.
No Protocol name entered.
Enter Protocol name.
110
WARNING
User attempted to access a MMC function,
but no MMC is inserted into the unit.
No Multimedia Card (MMC) inserted into unit.
Insert Multimedia Card (MMC) into unit.
111
WARNING
User attempted to upgrade software, but
the MMC in the unit is not a software
upgrade MMC.
Wrong Multimedia Card (MMC) inserted into unit.
Insert proper Mutimedia Card (MMC).
112
WARNING
Error upgrading Control Board software.
Unknown
Contact Dealer or Chattanooga Group for Service.
113
WARNING
Error upgrading motor board software.
Unknown
Contact Dealer or Chattanooga Group for Service.
114
WARNING
User pressed START, but has not yet set
the Max Level value to a value greater
than zero.
User has not completed Treatment Set Up.
Complete Treatment Set Up prior to pressing START.
115
WARNING
User pressed HOME button from treatment
review screen while a treatment was
running.
Home button has been pressed.
Press STOP to end treatment prior to pressing HOME.
Содержание TRITON 2841
Страница 15: ...13 Triton Traction Unit NOTES ...
Страница 72: ...Triton Traction Unit 70 9 SCHEMATICS Control Board 47801 1 of 3 ...
Страница 73: ...Triton Traction Unit 71 9 SCHEMATICS Control Board 47801 2 of 3 ...
Страница 74: ...Triton Traction Unit 72 9 SCHEMATICS Control Board 47801 3 of 3 ...
Страница 75: ...Triton Traction Unit 73 9 SCHEMATICS Motor Control Board 47803 1 of 2 ...
Страница 76: ...Triton Traction Unit 74 9 SCHEMATICS Motor Control Board 47803 2 of 2 ...
Страница 77: ...Triton Traction Unit 75 9 SCHEMATICS Patient Interrupt Switch 48056 1 of 1 ...
Страница 78: ...Triton Traction Unit 76 9 SCHEMATICS Serial Interface 47022 1 of 1 ...
Страница 79: ...Triton Traction Unit 77 9 SCHEMATICS Smart Card 47807 1 of 1 ...
Страница 80: ...Triton Traction Unit 78 9 SCHEMATICS sEMG Module 48123 1 of 1 ...
Страница 81: ...Triton Traction Unit 79 9 SCHEMATICS Block Diagram 1 of 1 ...
Страница 82: ...Triton Traction Unit 80 10 SPECIAL FIXTURES Calibration Fixture 1 of 4 ...
Страница 83: ...Triton Traction Unit 81 10 SPECIAL FIXTURES Calibration Fixture 2 of 4 ...
Страница 84: ...Triton Traction Unit 82 10 SPECIAL FIXTURES Calibration Fixture 3 of 4 ...
Страница 85: ...Triton Traction Unit 83 10 SPECIAL FIXTURES Calibration Fixture 4 of 4 ...