4

Safety

Product Manual - CryoDoser FleX

 LN

2

 Dosing System

Caution!  Before removing parts or 

loosening fittings, empty the 

CryoDoser FleX system of liquid 

and release any vapor pressure 

in a safe manner.

Safety Bulletin

Portions of the following information are extracted from 

Safety Bulletin SB-2 from the Compressed Gas Association, 

Inc  at 

www cganet com

. Additional information on oxygen, 

nitrogen, and cryogenics is available in CGA Pamphlet 

P-9. Write to the Compressed Gas Association, Inc., 1235 

Jefferson Davis Highway, Arlington, VA 22202..

Oxygen Deficient Atmospheres

Warning!  Nitrogen vapors in air may dilute 

the concentration of oxygen 

necessary to support or sustain 

life.

The normal oxygen content of air is approximately 21%. 

Depletion of oxygen content in air, either by combustion 

or by displacement with inert gas, is a potential hazard and 

users should exercise suitable precautions.

One aspect of this possible hazard is the response of humans 

when exposed to an atmosphere containing only 8 to 12% 

oxygen. In this environment, unconsciousness can be 

immediate with virtually no warning.

When the oxygen content of air is reduced to approximately 

15 to 16%, the flame of ordinary combustible materials, 

including those commonly used as fuel for heat or light, 

may be extinguished. Somewhat below this concentration, 

an individual breathing the air is mentally incapable of 

diagnosing the situation because the onset of symptoms 

such as sleepiness, fatigue, lassitude, loss of coordination, 

errors in judgment and confusion can be masked by a state of 

“euphoria,” leaving the victim with a false sense of security 

and well being 

Human exposure to atmosphere containing 12% or less 

oxygen leads to rapid unconsciousness. Unconsciousness can 

occur so rapidly that the user is rendered essentially helpless.  

This can occur if the condition is reached by an immediate 

change of environment, or through the gradual depletion of 

oxygen 

Most individuals working in or around oxygen deficient 

atmospheres rely on the “buddy system” for protection. 

Obviously the “buddy” is equally susceptible to asphyxiation 

if he or she enters the area to assist the unconscious partner 

unless equipped with a portable air supply. Best protection 

is obtainable by equipping all individuals with a portable 

supply of respirable air. Life lines are acceptable only if the 

area is essentially free of obstructions and individuals can 

assist one another without constraint.

If an oxygen deficient atmosphere is suspected or known to 

exist:

1  

Use the “buddy system.”  Use more than one “buddy” if 

necessary to move a fellow worker in an emergency.

2  

Both the worker and “buddy” should be equipped with 

self-contained or airline breathing equipment. 

Nitrogen

Nitrogen (an inert gas) is a simple asphyxiate. It will not 

support or sustain life and can produce immediate hazardous 

conditions through the displacement of oxygen. Under high 

pressure this gas may produce unconsciousness even though 

an adequate oxygen supply sufficient for life is present.

Nitrogen vapors in air dilute the concentration of oxygen 

necessary to support or sustain life. Inhalation of high 

concentrations of this gas can cause anoxia, resulting in 

dizziness, nausea, vomiting, or unconsciousness and possibly 

death. Individuals should be prohibited from entering areas 

where the oxygen content is below 19% unless equipped 

with a self-contained breathing apparatus. Unconsciousness 

and death may occur with virtually no warning if the oxygen 

concentration is below approximately 8%. Contact with cold 

nitrogen gas or liquid can cause cryogenic (extreme low 

temperature) burns and freeze body tissue.

Persons suffering from lack of oxygen should be 

immediately moved to areas with normal atmospheres.  

SELF-CONTAINED BREATHING APPARATUS MAY BE 

REQUIRED TO PREVENT ASPHYXIATION OF RESCUE 

WORKERS. Assisted respiration and supplemental oxygen 

should be given if the victim is not breathing. If cryogenic 

liquid or cold boil-off gas contacts worker’s skin or eyes, 

the affected tissue should be flooded or soaked with tepid 

water (105-115ºF or 41-46ºC). DO NOT USE HOT WATER.  

Cryogenic burns that result in blistering or deeper tissue 

freezing should be examined promptly by a physician.

Содержание CryoDoser FleX

Страница 1: ...Part Number 15960 Rev A 2017 Chart Inc Designed and Built by Chart Inc 46441 Landing Parkway Fremont CA 94538 USA 800 371 3303 Product Manual CryoDoser FleX Liquid Nitrogen Dosing System ...

Страница 2: ......

Страница 3: ... the Nozzle 7 Positioning the Dosing Head 9 Installing the Container Detect Sensor 9 Installing the Timing Sensor 9 Installing the Encoder 10 Installing the FleX Controller 11 Completed Assembly 11 Daily Operating Procedures 13 Dura Cyl Liquid Cylinder Fed System 13 Dura Cyl Cylinder Change Out Procedure 13 Bulk Tank House Fed System 14 Purging with Gaseous Nitrogen 14 System Start Up 14 System Sh...

Страница 4: ...cription A 08 30 2017 Original Specifications 19 CryoDoser FleX Technical Specifications 19 CryoDoser FleX Components 20 Standard Components 20 Optional Components 20 Common Replacement Parts 21 Sensor Positioning 22 Craft Custom Control Box Schematic 23 Craft Custom Interface Box Schematic 24 Warranty 25 ...

Страница 5: ...ssary for this function Key Benefits Lightweight PET reduce the weight of PET for cost and environmental savings Glass to PET Transition eliminate glass safety hazards and weight of containers Container Rigidity maintain container shape even with lighter weight containers Eliminate Paneling increase the internal pressure to offset paneling issues Oxygen Reduction create an inert environment to pre...

Страница 6: ...ponent damage Note A statement that contains information that is important enough to emphasize or repeat Acronyms Abbreviations The following acronyms abbreviations are used throughout this manual BAR Pressure Metric CPM Containers Per Minute GN2 Gaseous Nitrogen HMI Human Machine Interface Kg Kilogram LN2 Liquid Nitrogen mS Milliseconds O2 Oxygen PET Polyethylene Terephthalate PLC Programmable Lo...

Страница 7: ...d Personnel must wear protective gloves and eye protection whenever removing parts or loosening fittings Failure to do so may result in personal injury due to the extreme cold and pressure in the system Warning Accidental contact of liquid gases with skin or eyes may cause a freezing injury similar to a burn Handle liquid so that it will not splash or spill Protect your eyes and cover skin where t...

Страница 8: ... the buddy system for protection Obviously the buddy is equally susceptible to asphyxiation if he or she enters the area to assist the unconscious partner unless equipped with a portable air supply Best protection is obtainable by equipping all individuals with a portable supply of respirable air Life lines are acceptable only if the area is essentially free of obstructions and individuals can ass...

Страница 9: ... sure to carry with care Take care not to run into walls or drag the CryoDoser FleX unit on the ground If not installed immediately the CryoDoser FleX system should be stored in a location that is out of the way of frequent traffic and will prevent dirt water or other debris from getting inside the system Chart recommends storing the system in the crate when not in service Installation Application...

Страница 10: ...nce from the center of the support stand to the farthest point on the dosing head 4 Mark the location of the stand and install the four 5 8 bolts included with the support stand in the proper location Mounting the Dosing Body Image 3 depicts the dosing body with various parts not included for clarity The dosing body will come completely assembled on the bracket Loosen the 8 bolts on the black Delr...

Страница 11: ... the control box to this connector B Profinet Connector Connect the green Ethernet cable from the control box to this connector C Line Speed Connector For systems that will be using either an encoder or speed sensor connect the encoder or speed sensor cable to this connector D Timing Connector All systems will use this connector to detect a container or container pocket all Craft Custom users conn...

Страница 12: ...Re apply the dosing head heater Caution Never use any sharp or pointed object such as an ice pick screwdriver torch or similar device on the dosing head Image 10 The ribs of the internal bellows are a thin walled metal and the hole on the outer ring of the dosing head is a positive pressure port to help keep moisture out and ice from forming High heat and puncture holes will destroy the vacuum ins...

Страница 13: ... sensor with the CryoDoser FleX system Image 13 Ideal sensor placement is about four to six pockets from the dosing head If the sensor does not detect a container for the user defined number of seconds the CryoDoser FleX unit will stop dosing See the Inputs Outputs and Container Sensor sections for additional information The diagram below shows three different ways of positioning the sensor Instal...

Страница 14: ...functionality shielded cabling must be used The encoder should be grounded either thru the mounting bracket or separate ground wire Encoder Pin No Function Wire Color 1 Not Used N A 2 Not Used N A 3 Not Used N A 4 CH B Output Pink 5 Not Used N A 6 Not Used N A 7 Return 24Vdc Blue 8 Supply Voltage 24Vdc Red Shield The encoder location should be accessible for installation and maintenance The locati...

Страница 15: ...witch Main power switch for the FleX controller Dose Disable Button Manual dose enable disable button to disable dosing Light Tower Status The indicator light provides a visual display of the operating status of the CryoDoser FleX system Green Normal operation Amber solid Warning condition see Alarms section for additional information Red Fault condition see Alarms section for additional informati...

Страница 16: ......

Страница 17: ...levels it must be swapped with a full Dura Cyl cylinder Chart recommends a cylinder pressure of 60 psi and above 1 Close the liquid valve clockwise direction on the Dura Cyl cylinder 2 Disconnect the C Flex hose from the Dura Cyl cylinder using a 7 8 open end wrench or adjustable crescent wrench 3 Connect the C Flex hose to the liquid outlet on the full Dura Cyl cylinder using a 7 8 open end wrenc...

Страница 18: ...deep inside the vessel With the CryoDoser FleX system Chart has taken steps to ensure freeze ups are minimized Each system is provided with integral purging capabilities including an option to provide heated gas Heating the gas accelerates the removal of ice so as to bring the down time to a minimum For additional information about purging please refer to the Controller Quick Start Guide Caution W...

Страница 19: ...ad area b If the system will not be used for a short period of time it is helpful to keep some liquid in the dosing body to provide positive pressure at the vent preventing moisture from creeping in Nozzle use Standard nozzles are single opening devices with diameters ranging from 0 020 to 0 110 in 0 005 increments In addition to these devices Chart provides a range of specialty nozzles in our Sof...

Страница 20: ...igh speed dosing is about 500 pulses per container this encoder is capable of many more than needed to perform optimally How much LN2 is needed Identifying the right amount of liquid can be challenging Some trial and error is commonplace However there are a couple of ways to minimize the number of trial and error situations 1 Use the largest nozzle and lower the dose duration It is much easier to ...

Страница 21: ...dosing head heater is 150 F 65 C Caution If possible always perform nozzle change out procedures before introducing LN2 into the CryoDoser FleX unit Failure to do so may cause the nozzle to unthread and fall out Note If the nozzle does not loosen easily a low voltage heat gun may be used to heat the end of the nozzle tool The tool can then be placed over the nozzle to slowly melt any ice allowing ...

Страница 22: ...m at 1 408 371 4932 for a detailed procedure No liquid from the dosing head There is insufficient liquid inside the CryoDoser FleX unit The unit is disabled The nozzle is frozen shut The container detect sensor is not detecting a container Check the level of LN2 If the level is empty or low open the supply valve Pull out the Dose Enable switch to enable mode ensure switch is not lit Run Insta Purg...

Страница 23: ...50 8mm Arm Reach Short 20 1 2 Long 27 1 2 20 1 2 to 27 1 2 521mm to 699mm Head Pressure Variable Nozzles Ships with 0 040 0 050 0 060 0 020 0 100 available 0 005 increments System Utilities Liquid Nitrogen 3 120 psi 0 2 8 3 bar Gaseous Nitrogen 60 80 psi 4 1 5 5 bar Electricity 110 240V AC 50 60Hz 110W Steady State Consumption 0 04 gal 0 15 liter hour Purge Feature Yes Vacuum Insulated Yes Arm Yes...

Страница 24: ...d The dosing head delivers the dose of LN2 Dosing Head Heater The CryoDoser FleX unit has a self regulating dosing head heater The maximum temperature of the dosing head heater is 150 F 65 C and prevents frost or ice formation at the dosing head area The heater is held in place by a set of o rings If needed the dosing head heater can be removed by slipping it off of the dosing head The dosing head...

Страница 25: ...ming 35067 004 12 Photoelectric Sensor Assy Container 35067 003 13 Encoder Assy 35067 002 14 Control Box Assy CC 35065 14a Control Box Assy PP 35063 15 Interface Box Assy CC 35064 15a Interface Box Assy PP 35066 16 Doser Stand 119K 17 Doser Base 326A 18 Y Pattern Valve S6OMYCD 19 123 Fill Hose 123 S 20 10 Micron Filter CR_108 21 AC Power Cord 410 22 Ethernet Cable 2840 122 23 DC Power Interconnect...

Страница 26: ...22 Specifications Product Manual CryoDoser FleX LN2 Dosing System Sensor Positioning ...

Страница 27: ...23 Specifications Product Manual CryoDoser FleX LN2 Dosing System Craft Custom Control Box Schematic ...

Страница 28: ...24 Specifications Product Manual CryoDoser FleX LN2 Dosing System Craft Custom Interface Box Schematic ...

Страница 29: ...n thirty 30 days after receipt of the LN2 Dosing Systems Chart shall inspect the returned LN2 Dosing Systems and if said LN2 Dosing Systems is found defective shall at Chart s option as purchaser s sole and exclusive remedy either i repair Warranty or replace such LN2 Dosing Systems or any defective component or part thereof which proves to be defective or ii refund the net purchase price paid by ...

Страница 30: ......

Отзывы: