Pulse Oximeter

   

ACCURO 

User Guide

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Charmcare Co., Ltd. 

 

Pulse Oximeter ACCURO                                                                                                    www.charmcare.com 

 

-2-                                                          OP-EN-06 

 

Содержание ACCURO

Страница 1: ...Pulse Oximeter ACCURO User Guide Charmcare Co Ltd Pulse Oximeter ACCURO www charmcare com 2 OP EN 06 ...

Страница 2: ...he Power and Grounding the Product 16 Installing the Stand 17 Part 3 Using the Product 19 Preparations Before Use 19 Pulse Oximeter ACCURO www charmcare com 4 OP EN 06 Operations 20 The Screen 21 SpO2 PR Data Area 22 System Status Area 24 Using the Buttons and Alarm Lamp 25 Using Power 27 Part 4 Measuring the SpO2 28 Attaching the SpO2 Probe 28 Measuring the SpO2 30 Part 5 Swich Screen Mode 31 Swi...

Страница 3: ...tric patients via finger in hospitals medical facilities and subacute environments Charmcare Tabletop Pulse Oximeter is intended for spot checking and or continuous monitoring of patients the sensor of device is reusable or disposable To obtain information about a warranty if any for this product contact your local Charm care representatives This product has been manufactured and inspected followi...

Страница 4: ...ed service or repairs Charmcare Co Ltd Tel 82 2 862 5052 Fax 82 2 862 5065 Need technical advice Charmcare Co Ltd Tel 82 2 862 5052 Fax 82 2 862 5065 Website HTTP www charmcare com Contact Us Battery Replacement Pulse Oximeter ACCURO www charmcare com 8 OP EN 06 Certain parts of this manual have been highlighted Users are advised to pay close attention to warnings and cautions Manufacturer and or ...

Страница 5: ...the Product Pulse Oximeter ACCURO www charmcare com 10 OP EN 06 Warning Do not use the product in high ambient light conditions direct sunlight Please check the following conditions before attempting to use the product Are you using the proper power source line 100 240VAC Are you using the power source supplied by the manufacturer DC 12V Are all parts power cord and optional parts connected proper...

Страница 6: ...ution Do not dispose single use probe to any hazard place Always think about environmental contamination Maintenance and Cleaning Pulse Oximeter ACCURO www charmcare com 12 OP EN 06 After use store the product in an area that does not exceed 85 humidity and maintain the temperature in the range of 0 C 40 C Do not store the product together with chemical products and or in an area subject to gas ex...

Страница 7: ...the device to suit your needs and patient status results may be printed out on a printer Basic Accessories Type Quantity SpO2 Sensor ACCY 0A0PRB 1 DC Power Unit 0 Power Cord 1 User s Manual 1 Optional Parts Battery BTLI503465L9022S1P Printer Printer Paper Pole Bracket Parts Pulse Oximeter ACCURO www charmcare com 14 OP EN 06 Front view 1 Probe Port 2 Display Screen 3 Buttons 4 Alarm Lamp 5 Using A...

Страница 8: ...is between 0 40 and the humidity is 85 Make sure that the power cord is plugged in properly Do not plug in more than one device into one outlet Ground the product if noise develops Do not use a power cord that causes noise The product is very sensitive to shock Caution is advised Keep the product dust free and install away from flammable materials Connecting the Power Connect the power to the powe...

Страница 9: ...irst decide how you want the product to be mounted Then insert two bolts into the holes on the stand and tighten REAR Fig Attaching the stand Installing the Stand Pulse Oximeter ACCURO www charmcare com 18 OP EN 06 Step 2 Mount the stand to the pole by sliding the stand onto the pole and then turn the handle to tighten Caution The pole on which the stand is mounted should be less than 30mm in diam...

Страница 10: ...ched on to the bridge of the nose or forehead When attaching the SpO2 sensor on to the finger no manicure should be on the finger nails For sterilization or antiseptic cleaning is necessary for the SpO2 select the methods shown below according to the environment conditions of the medical institution equipment and use the usage and application of the professional prefers Preparations Before Use Pul...

Страница 11: ...06 NO Name Description 1 Oxygen Saturation SpO2 reading Displays the measured value of SpO2 When SpO2 exceeds the alarm limit the value will start blinking 2 Spo2 low alarm limit Displays the setup low alarm limit When the SpO2 is less than this value the value will turn red 3 Spo2 High alarm limit Displays the current set SpO2 High alarm limit value When the SpO2 value is larger or the same the v...

Страница 12: ...ion 1 Battery Status Displays the battery status Uses AC power and battery is in charge Uses a battery and capacity is FULL Uses a battery and capacity is 3 4 Uses a battery and capacity is 1 2 Uses a battery and capacity is 1 4 Uses a battery and capacity is low 2 Alarm sound Status Displays the alarm souund off state Displays the alarm sound paused state and left time Displays the Alarm sound of...

Страница 13: ...p in the general screen 6 button Increases the selected value in the menu screen Using the Buttons and Alarm Lamp 1 2 3 5 9 7 6 4 8 Pulse Oximeter ACCURO www charmcare com 26 OP EN 06 Changes the screen brightness in the general screen 7 MENU button Moves to the SETUP menu when pressed in the general screen 8 Power Used when turning on off the power of the product Press the power for more than 2 s...

Страница 14: ...hows in the lower part of the screen saved data can be damaged Service request is necessary Using Power 1 2 Pulse Oximeter ACCURO www charmcare com 28 OP EN 06 Part 4 Measuring the SpO2 Step 1 Wipe down the area where you wish to place the probe with alcohol Step 2 Attach the probe to the patient s finger Step 3 To get an accurate reading make sure the patient minimizes all movement and Please att...

Страница 15: ...tients who have high fevers or have problems with distal circulation will be 2 3 degrees higher than normal Patients with abnormally high oxyhemoglobin or methemoglobin levels will not give a proper SpO2 reading Caution Taking the NIBP can affect the SpO2 reading When taking NIBP place the SpO2 probe on the other arm Avoid using the probe with other medical equipment that affects blood flow Avoid ...

Страница 16: ...II consists of 4 types of screens and the can be switch to the screen the user designates Press the button once to switch the screen in the order of SCREEN1 SCREEN2 SCREEN3 SCREEN4 SCREEN 1 SCREEN 2 Switch Screen Mode Pulse Oximeter ACCURO www charmcare com 32 OP EN 06 SCREEN 3 SCREEN 4 ...

Страница 17: ... menu using the button and button Step 2 Setup the value by using the button and button in the designated menu Step 3 Press the MENU menu button from EXIT to move to the SETUP menu The contents of each item are as shown as below Item Description Range PULSE VOLUME Set the degree of pulse volume 0 7 BRIGHTNESS Set the brightness of the screen 1 5 SCREEN TYPE Set the main screen mode Part 6 Transfer...

Страница 18: ...d button Step 2 Adjust the setting value using the buttons according to the selected menu Step 3 Press the MENU menu button from EXIT to move to the SETUP menu The contents of each item are as shown as below Item Description Range ALARM VOLUME Set the degree of the alarm volume 0 7 ALARM DURATION Set the duration second when in alarm sound pause and alarm sound off 30 60 90 120 ALARM 0 REMIND When...

Страница 19: ...cks if the PR measurement value is out of the range of the setup range during alarm watch PR alarm watch status displays when setting is OFF OFF ON Setup ALARM Limit Pulse Oximeter ACCURO www charmcare com 38 OP EN 06 SPO2 ALARM LEV Set the level of the SPO2 ALARM LOW MEDIUM HIGH PR ALARM LEV Set the level of the PR ALARM LOW MEDIUM HIGH FINGER OFF LEV Set the alarm level when FINGER OFF occurs LO...

Страница 20: ...LT PEDIATRIC NEONATE When the Patient Mode changes the alarm limit value and SPO2 average switches to the default value specified in Attachment 1 Setup PATIENT Pulse Oximeter ACCURO www charmcare com 40 OP EN 06 Step 1 Move to the designated menu using the button and button Step 2 Adjust the setting value using the buttons according to the selected menu Step 3 Press the MENU menu button from EXIT ...

Страница 21: ...he main screen Check TREND Screen Save Measured Data Pulse Oximeter ACCURO www charmcare com 42 OP EN 06 2 View Tabular Trend Screen Moves to the Tabular Trend screen by selecting SETUP MENU TREND TABULAR as shown below Displays by the recent data order No Description 1 Button key instructions Upper button Displays the previous page Moves quickly to the page after pressing more than 2 seconds Lowe...

Страница 22: ...on to move to the previous page 1 7 8 9 10 5 3 2 4 6 Pulse Oximeter ACCURO www charmcare com 44 OP EN 06 4 Displays the status whether moving to the next page is possible Press the button to move to the previous page 5 The trend bar of 6 moves using the button and button 6 Displays the location of the trend bar Can move to check to the designated area like 4 7 Displays the saved date and time in t...

Страница 23: ...ient Medium Priority Occurs when the physiological signal is abnormal or requires treatment Low Priority Occurs when the user has to be notified or the system confronts an error during Alarm Levels Alarm categories Pulse Oximeter ACCURO www charmcare com 46 OP EN 06 measurement Patient alarm and Technical alarm are prioritized according to the content and some parts can be set by the user from ala...

Страница 24: ...d in alarm messages and alarm sound operations for limit alarms The limit alarm watch of SpO2 can stop when SETUP ALARM SPO2 ALARM is unchecked And the alarm watch status switches to in the SpO2 screen window To stop the limit alarm watch of PR uncheck the SETUP ALARM PR ALARM And the alarm watch status switches to in the PR screen window Alarm sound off This turns the alarm souund off It operates...

Страница 25: ...URO www charmcare com 50 OP EN 06 Verifying Visual and Audible Alarm Indication If the monitor fails to perform as specified in this test contact qualified service personnel or your local supplier for assistance You can verify the alarm operation for all parameters SpO2 by following the below procedures 1 Connect the monitor to an AC power source 2 Press Power button to turn on the monitor 3 Conne...

Страница 26: ...Connecting the Printer Pulse Oximeter ACCURO www charmcare com 52 OP EN 06 Printing The printer allows you to print the patient measurements for the designated trend Step 1 Move to the screen by SETUP TREND TABULAR Step 2 Press the button on the TRAND screen to start the printing The printing status will be shown on the print status display part on the lower portion of the screen ...

Страница 27: ...w how to proceed Try the following steps to work through the problems General Situation Cause The power will not come on Make sure the product is plugged in Make sure the battery is fully charged I can t hear the alarm or the pulse sounds Make sure the alarm is not turned off Coin battery image is shown in the lower portion of the screen Make sure the coin battery is not lacking capacity The saved...

Страница 28: ...ut 100 240 Vac 50 60Hz Fuse rating 5A 250V Pulse Oximeter ACCURO www charmcare com 56 OP EN 06 Battery Type Li ION internal battery Charge time 5hours Battery Capacity Typically 5hours using new fully charged battery Warranty 6 months Environmental Operating Temperature 5 40 Transport Storage Temperature 20 70 Operating Humidity 15 95 noncondensing Operating Attitude 1 000 ft 12 000 ft Physical Ch...

Страница 29: ...ds Memory Save continuously for 15 days 10 seconds unit Display Tabular Graphic 1 72 hour unit External Print Serial RS 232 Connect to exterior printer or PC Digi View Pulse Oximeter ACCURO www charmcare com 58 OP EN 06 Attachment 1 Default Value The value maintains after setting the value of the oxygen saturation device and turning the power off and back on The default value of the product is as ...

Страница 30: ... set to a fixed value Patient Alarm Message Alarm Message Level Cause ASYSTOLE High Heartbeat is not sensed Check the status of the patient HIGH SPO2 Medium Occurs when the measured value is greater than the high limit or greater than the low limit Check the status of the patient Also check if the limit set value is valid LOW SPO2 Medium HIGH PR Medium LOW PR Medium Technical Alarm Message Alarm M...

Страница 31: ...armcare com 61 OP EN 06 Charmcare Co Ltd 714ho Woolim Lions2 cha Gasan Dong Geumcheon Gu Seoul Korea ZIP 153 787 Tel 82 2 862 5052 Fax 82 2 862 5065 Home Page HTTP www charmcare com Product ACCURO E Manual Version 1 04 2012 09 12 ...