Calibration Unit for Testers

Installation, Operation and Maintenance

Description

The Charleswater 

99090

 Calibration Unit is designed to 

verify whether a tester is operating within specifications.  

The 

99090

 Calibration Unit is a passive device and 

requires no power source.  The Calibration Unit is 

manufactured with industry accepted test ranges for 

both wrist straps and foot grounders.  Both the wrist 

strap and footwear pass range is set at 750K - 35M.  

The Charleswater Calibration Unit is calibrated to NIST 

traceable standards. 
The Charleswater 

99090

 Calibration Unit is to be used 

with the following items:

Item

Description

99037

Combo Tester

99038

Combo Tester with Stand and Foot Plate

Packaging

1  Calibration Unit

2  Test Leads with Banana Plug Terminals

1  Certificate of Calibration

August 2015

Figure 1.  Charleswater 

99090

 Calibration Unit

Operation

COMBO TESTER
Testing the Wrist Strap Circuit

1. 

Plug one of the included test leads into the black 

banana jack labeled “E” on the Calibration Unit.  

Connect the opposite end of the test lead into the 

yellow banana jack labeled “WRIST STRAP” on the 

Combo Tester.  Toggle the switch to the “WRIST 

STRAP” position.

  TECHNICAL BULLETIN PPE-5034.E

Figure 2.  Connecting the test lead from banana jack “E” 

on the Calibration Unit to the WRIST STRAP banana 

jack on the Combo Tester

2. 

Connect the second test lead to the red 675K 

banana jack labeled “A” on the Calibration Unit.  

Touch the opposite end of the test lead to test plate 

on the Combo Tester and press down to activate 

the test.  Be sure not to touch the banana plug or 

test plate with your skin.

3.  Observe the response from the Combo Tester.  The 

alarm should sound, and the LOW LED should 

illuminate red.

4. 

Repeat the procedure for test points B (825K), C 

(35M) and D (40M).  The expected test results can 

be found in the following table.

Test 

Point

Resistance 

Value

Test Output

A

675K

FAIL LOW

B

825K

PASS

C

35M

PASS

D

40M

FAIL HIGH