1
USER MANUAL
MS2400
Infant Scale
Please keep the instruction manual at hand all the time for future reference.
Страница 1: ...1 USER MANUAL MS2400 Infant Scale Please keep the instruction manual at hand all the time for future reference...
Страница 2: ...for 2019 Separate collection for waste of electrical and electronic equipment in accordance with Directive 2002 96 EC Indicates that device conforms to 93 42 EEC as amended by 2007 47 EC Medical Devi...
Страница 3: ...rnational copyright law All content is licensed and usage is subject to written authorization from Charder Electronic Co Ltd hereinafter Charder Charder is not liable for any damage caused by a failur...
Страница 4: ...12 A Height Measure Attachment 12 B Inserting Batteries 14 C Using AC Adapter 14 III Indicator 15 IV Using Device 16 A Basic Operation 16 B Tare 16 C Recall 16 V Device Setup 17 VI Troubleshooting 19...
Страница 5: ...y surface Usage on soft surfaces ex carpet may result in inaccurate results Ensure all parts are properly locked and tightened before operating the device Safety Instructions Before putting device int...
Страница 6: ...routine maintenance However regular checking of accuracy is recommended frequency to be determined by level of use and state of device If results are inaccurate please contact local distributor Warran...
Страница 7: ...able and avoid sharp edges Do not overload extension cables connected to the device Route cables carefully to avoid tripping Keep device away from liquids Do not remove the plug by yanking on the cabl...
Страница 8: ...dance RF emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic e...
Страница 9: ...ital environment Surge IEC 61000 4 5 1kV line s to line s 2kV line s to earth 1kV line s to line s 2kV line s to earth Mains power quality should be that of a typical commercial or hospital environmen...
Страница 10: ...AM at 1 kHz 3 V m 80MHz to 2 7 GHz Portable and mobile RF communications equipment should be used no closer to any part of the device including cables than the recommended separation distance calcula...
Страница 11: ...he MS2400 Infant Scale The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the device can help prevent electro...
Страница 12: ...Height Measure Attachment 1 Attach bracket to device and fasten screws using screwdriver Version with rubber head pad Version without rubber head pad Attach bracket here rubber head pad rubber head pa...
Страница 13: ...sure attachment HM80D or HM80M to bracket A clicking noise will be heard 3 To detach height measure find the latch located at the rear of the attachment Press down on the buckle and gently remove the...
Страница 14: ...ery cover Step 2 Remove battery cover Step 3 Insert 9V battery Step 4 Replace battery cover C Using AC Adapter Device can be powered using AC adapter located at rear of device Plug adapter into device...
Страница 15: ...ons 1 RECALL Recall and release weight stored in memory 2 TARE Deduct weight from reading after measurement 3 ON ZERO OFF When device is off press to power on When device is on press to zero weight un...
Страница 16: ...rload B Tare The tare function allows the user to deduct the weight of objects from the device s measurement result Tare can be used when weight of object is 400g at above 2 of full 20 kg capacity 1 P...
Страница 17: ...ter submenu Auto Power Off Instruct device to shut off automatically after a certain period of time Auto off options 120 sec 180 sec 240 sec 300 sec off Press TARE to toggle between time options and R...
Страница 18: ...i Fi function can be turned on or off Press TARE to toggle between on off and RECALL key to confirm selection Wi Fi Setting optional If device has Wi Fi module installed this option will appear Press...
Страница 19: ...l or electrical interference unless damage is attributable to negligence on the part of Charder If device is not covered under warranty a service maintenance charge will apply plus cost of replacement...
Страница 20: ...or replacement services 1 Device will not power on Faulty on off key Broken or damaged wires causing short circuit or faulty connection Safety fuse burnout Faulty adapter 2 Indicator damage Possible h...
Страница 21: ...asurement platform and try again Over Zero Zero count is higher than setting range Device will turn off automatically in 30 seconds if status continues Error normally caused by faulty loadcell or wiri...
Страница 22: ...LCD screen 4 1 2 digits Dimensions Device plus Handle 640 W x 355 D x 60 H mm Platform Surface 500 W x 350 D mm Device Weight 2 8 kg Key Functions On Zero Off Tare Recall Power Supply 9V battery adapt...
Страница 23: ...12V 2A AD 8057 AD 0520 UE24WU 120200SPA EU 12V 2A AD 8056 AD 0519 UE24WU 120200SPA UK 12V 2A AD 8074 AD 0534 UE24WU 120200SPA AU 12V 1A AD 8095 UE24WCP1 120100SPA US 12V 1A AD 8095 UE24WCP1 120100SPA...
Страница 24: ...d directives 93 42 EEC as amended by 2007 47 EC Medical Device Directive 2014 31 EU Non automatic Weighing Instruments Directive Please see separate document showing on sticker of device for above CE...