SECTION 2
–
OPERATING INSTRUCTIONS 2-5
Cooling the Cylinder
1.
Set the Temperature Controller to Manual mode by pressing the F1 button. The
“
hand
”
icon will appear.
2.
Turn the HEATER SWITCH (#14
) to “OFF.”
3.
Turn the CYLINDER COOLING SWITCH (#8
) to “ON.”
4.
The internal cooling coils provide cylinder cooling with resulting reduction of pressure
due to thermal contraction. The control limits on the pressure gauge will maintain
pressure through the cooling cycle.
5.
Continue until the cylinder temperature is reduced as desired.
Caution: When the cooling water is turned off, heat from the cylinder will cause the
temperature to rise.
Emptying the Cylinder
Note that the instrument is designed to drain the mineral oil from the cylinder to the waste oil
port at the back of the instrument. This allows contaminated oil to be safely removed for
disposal or recycling. If uncontaminated oil must be returned to the internal reservoir,
connect the waste oil port to the port located at the top of the internal oil reservoir.
Caution: The waste oil may be hot from a test
1.
Turn “ON” the AIR EXHAUST SWITCH (#
9). Wait for the pressure to decrease to
zero.
Содержание AMETEK 5617
Страница 3: ......
Страница 9: ...P 4 PREFACE This page is intentionally left blank ...
Страница 23: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 29: ...5 6 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Страница 31: ...6 2 SECTION 6 DRAWINGS AND SCHEMATICS This page is intentionally left blank ...
Страница 32: ......
Страница 33: ...CHANDLER ENGINEERING ...
Страница 36: ...TB1 TB2 ...
Страница 37: ...CHANDLER ENGINEERING ...
Страница 38: ......
Страница 41: ...cut out postcards on dotted lines ...