Chapter V Trigger System
28
Fig. 5.11 Pulse Trigger Example
Runt trigger
It will be triggered when the input signal crosses through a threshold but fails to cross through the second
threshold before crossing through the first threshold again.
High
Threshold
Negative Runt
Low
Threshold
Positive Runt
Fig. 5.12 Schematic Diagram of Runt Pulse
Trigger type
Press down
【
Trigger
】
key of the trigger control area on the front panel, and open the trigger menu.
Click [Type] menu key and click [Runt] key in the popup Type menu to select the Runt trigger.
Fig. 5.13 Runt Trigger Menu
The relevant Runt trigger information
will be displayed on the top right corner of the screen,
including the source and polarity.
Source:
Click [Source] menu key, and the Source selection menu will pop up on the right side of the screen. The CH1
CH4 can be made as the source.