EN
OPERATOR'S MANUAL
49
6.2. POSITIONING THE PATIENT FOR 3D EXAMINATIONS
Only for 3D machines
1) After preparing the X-ray device, let the patient in.
2) Adjust the height of the motorised column with the buttons
to facilitate patient access. The column
must reach the height of the patient.
3) The patient must grip the handles with both hands and
stand straight.
4) During 3D examinations, the patient's head must be
positioned in one of these 3 positions: right-hand,
central or left-hand. The right-hand or left-hand
position is fundamental to centre the field of view in
the lateral anatomical districts. Such positioning is
reached by manually moving chinrest and craniostat
in the desired direction.
A position detector on the chinrest will indicate whether the chosen position is correct for the selected
examination. If it is incorrect, the software will not allow to continue the examination until a correct position is
reached.
5) Use the lateral laser guides to obtain the desired
patient alignment.
6) Place the patient's head using the frontal laser beam as
a guide since it identifies the sagittal plane passing
through the centre of the selected FOV.
7) Go to the PC workstation.
6.3. PERFORMING THE EXAMINATION
Follow the instructions for the acquisition operations described in the NNT manual to complete image acquisition.
Refer to “Extended View” mode for the acquisition with fields of the Extended View type.
Only for 3D machines
Follow the instructions for the acquisition operations described in the NNT manual to complete image acquisition.