2 About safety
2.5.1 Positioning the animal
Ensure that the animal is correctly positioned on the table and no part of its body will bump into the device or
be squashed during the positioning stage.
Ensure that there is no possibility of any animal hair becoming caught.
Carry out the same checks for any catheters, respiratory tubes or ECG (electrocardiograph) cables.
Do not start any movement until you have ensured the safety of the animal and that there is nothing
obstructing the movement of the table itself.
For further details on positioning the animal, please refer to par. 6.1.2 - “Positioning the animal and starting a
new scan“.
2.5.2 During the scan
Please refer to the paragraph of the same name in the “User Manual”.
2.5.3 Removing the animal from the scanning area
At the end of the examination or after the emergency stop button has been pressed, the patient table can be
extracted from the scanning area and the animal removed.
2.5.4 Extracting the animal in the event of the breakdown / malfunction of the
patient table
In the event of an interruption to the operation of the patient table with stretcher, extract the animal by
manually removing the stretcher completely from the gantry of the device.
2-2
Veterinary Application
Содержание NewTom 5G Vet
Страница 1: ...0051 97050812 Rev 1 28 10 2016 ATTACHMENT to the User Manual Veterinary Application Attachment...
Страница 38: ...10 APPENDIX A Technical references 10 8 Veterinary Application...
Страница 39: ...10 APPENDIX A Technical references Veterinary Application 10 9...
Страница 40: ...10 APPENDIX A Technical references 10 10 Veterinary Application...
Страница 41: ...10 APPENDIX A Technical references Veterinary Application 10 11...