Patient Table Service Procedures
3-10
The following table includes error codes that could appear on the console display and possible causes /
solutions.
Err. Code
Description
Possible causes / Solutions
1
Power supply of Hanning boards < 26VDC
40VDC power supply of Hanning boards may be
missing or may not be connected properly
2
Power supply of Hanning boards > 43VDC
40VDC power supply of Hanning boards not be
connected properly.
5
Motor or Hanning pantograph board overheating
Wait 2 minutes and try again the movement. If the
error persists, there may be a malfunction of the
pantograph motor or the pantograph is not free to
move.
9
Overcurrent peak output to Hanning pantograph
board
There may be a malfunction of the pantograph motor
13
Short-curcuit output to Hanning boards
The pantograph motor may be faulty, or incorrect
wiring of the motor power supply: X102, X103, X104
on one of the Hanning boards.
16
Enable external command not present on Hanning
boards
May be missing the enable jumper between pins 3 and
2 of X208.
17
Hanning pantograph board movement sequence time
out
The Hanning board may be faulty or not programmed
properly or the pantograph motor is locked
18
Error on the start of the movement of the Hanning
pantograph board.
The Hanning board may be faulty or not programmed
properly
128
Internal timing error on one of the Hanning boards.
The Hanning board may be faulty or not programmed
properly
129
System error of Hanning boards
The Hanning board may be faulty or not programmed
properly
131
Reset for Hanning boards error
The Hanning board may be faulty or not programmed
properly
132
Reset for power supply change in Hanning boards
40VDC power supply of Hanning boards may have
undergone rapid > 5 VDC
133
Software reset of Hanning boards
The Hanning board may be faulty or not programmed
properly
140
CRC error on parameters reading of hanning boards
The Hanning board may be faulty or not programmed
properly
141
Read error of Hanning board table
The Hanning board may be faulty or not programmed
properly
142
Read error on parameters of Hanning boards
The Hanning board may be faulty or not programmed
properly
150
XY axes controller board doesn’t send regularly the
status message.
The board may be turned off, faulty or CANBUS is not
working properly.
151
Hanning pantograph board doesn’t respond correctly
if cyclically interrogated
The board may be faulty, not powered or with wrong
address. CAN may be faulty
152
The Hanning board that control the stretcher position
and handle release doesn't respond properly if
cyclically interrogated
The board may be faulty, not powered or with wrong
address. CAN may be faulty
210 ÷ 213
Error during motion started by remote control
The system can not reach the required positions or
the movement is stopped before its completition
253
There is a position mismatch between the
potentiometer and pantograph encoder
It may not have been carried out a correct
calibration, the Hanning board may have lost some
encoder count or may be an hardware problem to
encoder, potentiometer or wiring.
254
Calibration data stored in the console are not
compatible with the new firmware version just
installed and consequently restored
The console board may have been replaced or
updated with new firmware without calibrations
Содержание NewTom 5G Series
Страница 1: ...97050267 Rev 3 28 10 2016 NewTom 5G series Patient Table Service Manual EN...
Страница 21: ...NewTom 5G series Patient Table Service Procedures 3 13 CONSOLLE BOARD 97661095 CONSOLLE BOARD...
Страница 22: ...NewTom 5G series Patient Table Service Procedures 3 14 3 1 8 Screens flow chart...