 Patient Table Service Procedures 

 

3-6

 

The following table shows the meaning of each limit switch 

 

Limit switch 

L status 

H status 

Description 

Stretcher Longitudinal 

movement switches 

 

 

Pantograph all the way out of the gantry 

 

 

 

Pantograph all the way toward the gantry 

 

Stretcher Transversal 

movement switches 

 

 

Stretcher completely on the left 

 

 

 

Stretcher completely on the right  

 

Pantograpg vertical 

movement switches 

 

 

Pantograph in the lowest position 

 

 

 

Pantograph in the highest position 

 

LT 

Stretcher completely out 

of the gantry switch 

 

 

Stretcher all the way out of the gantry 

UT 

Unlock table switch 

 

 

Stretcher lock 

Anti-pitch safety switch 

● 

 

AntiPinch activated 

Emergency button switch 

● 

 

Emergency button activated 

 

3.1.3 

 

DATA DEBUG page 

 

In the 

DATA DEBUG page

 appear data about the system status. 

 

 

DATA DEBUG page 

 

REC=    

N°of received data packets with errors 

TEC=    

N°of trasmitted data packets with errors 

CAN CA0=  

CAN tree status: if  “Ok” connection is good, if “NOk” connection failed 

CAN CA1=  

CAN tree status: if  “Ok” connection is good, if “NOk” connection failed 

CAN CA2=  

CAN tree status: if  “Ok” connection is good, if “NOk” connection failed 

E0,E1,E2,E3=  Contains the last four error codes 

 

 
Press any button to return to the POS page. 

 
 

Содержание NewTom 5G Series

Страница 1: ...97050267 Rev 3 28 10 2016 NewTom 5G series Patient Table Service Manual EN...

Страница 2: ...269 EEC 90 270 EEC 90 394 EEC and 90 679 EEC concerning the improvement of the health and safety of workers in the workplace and with the essential requirements set forth by the Italian Legislative De...

Страница 3: ...4 2 1 3 DATA DEBUG page 2 4 3 PATIENT TABLE WITH STRETCHER 3 1 3 1 SERVICE LEVEL PROCEDURES 3 1 3 1 1 Service Level Access 3 1 3 1 2 TOOLS page 3 2 3 1 2 1 ACTION page 3 2 3 1 2 2 LIMIT SWITCH page 3...

Страница 4: ...please refer to the Service Manual document This manual is intended for trained personnel recognized by the manufacturer of the NewTom 5G series devices Prior to operating or servicing this device thi...

Страница 5: ...ject and or person 2 1 1 Service Level Access From the INFO page hold the button and press the button The SERVICE Level modality will be enabled In order to disable the SERVICE Level modality repeat t...

Страница 6: ...alue using the and buttons The value will be incremented decremented by 1 10 or 100 units depending on the pressure duration The value is reset every time the user scroll among different actions and i...

Страница 7: ...anical parts LOAD DEFAULT PW Restore the original manufacturer settings without resetting the calibrations GO PACKAGING PW Bring the table to the position for the packaging SET HORIZ POS PW Store in t...

Страница 8: ...gnals In the bottom right corner the software version is shown By pressing the MODE button is possible to return to the TOOLS page LIMIT SWITCH page X X Axis gantry in out Y Patient s left right Z Pan...

Страница 9: ...th object and or person WARNING Every movement can be stopped by pressing the emergency button located under the console keypad only if the patient table is powered through the 5G series gantry 3 1 1...

Страница 10: ...the F2 button POS page by pressing MODE button 3 1 2 1 ACTION page The ACTIONS page contain a scroll menu with a list of possible action to perform The F1 and F2 buttons allow to move among the list...

Страница 11: ...he and buttons Change digit by the arrows keys NEW TABLE POT TYPE To setup only in case the small potentiometer cod 97643500 is mounted enabled by default Perform only for patient table with S N 0005...

Страница 12: ...ress OK to confirm 5 Turn off the machine 6 Connect C6 connector 5G patient table J8 connector and X207 connector 7 Turn on the machine and perform the Pantograph potentiometer calibration by select P...

Страница 13: ...RDS command from the service menu insert the password and press OK to confirm 12 Turn off the machine 13 Connect C6 connector 5G patient table J8 connector and X207 connector 14 Turn on the machine EX...

Страница 14: ...completely out of the gantry switch Stretcher all the way out of the gantry UT Unlock table switch Stretcher lock A Anti pitch safety switch AntiPinch activated E Emergency button switch Emergency bu...

Страница 15: ...k the pantograph by using BACK button In case of table installation console replacement or firmware update perform the following calibrations 1 Insert the device serial number following procedure SET...

Страница 16: ...n case of potentiometer replacement or after adjustment of the limit switch position follow the next steps 1 Select the type of mounted potentiometer following procedure NEW TABLE POT TYPE OLD TABLE P...

Страница 17: ...e quota 2 mm Longitudinal pantograph movement relative quota 200 210 mm Vertical pantograph movement relative quota 300 315 internal units The following table shows the stretcher pantograph the correc...

Страница 18: ...boards error The Hanning board may be faulty or not programmed properly 132 Reset for power supply change in Hanning boards 40VDC power supply of Hanning boards may have undergone rapid 5 VDC 133 Sof...

Страница 19: ...NewTom 5G series Patient Table Service Procedures 3 11 3 1 6 Hanning boards connections and interfaces 97661177 97661178...

Страница 20: ...3 1 7 List of patient table electronic boards UPPER STRUCTURE BOARDS 1 97661178 97660848 TABLE BOARD 2 97660897 XY MOTORS CONTROL BOARD 3 97660888 MANAGEMENT BOARD PANTOGRAPH BOARDS 1 97660905 POWER S...

Страница 21: ...NewTom 5G series Patient Table Service Procedures 3 13 CONSOLLE BOARD 97661095 CONSOLLE BOARD...

Страница 22: ...NewTom 5G series Patient Table Service Procedures 3 14 3 1 8 Screens flow chart...

Страница 23: ...emark of CEFLA s c All other products and brand names are registered trademarks or trademarks of their respective companies NEWTOM 5G is manufactured by CEFLA s c Phone 39 045 8202727 Fax 39 045 82030...

Отзывы: