OPERATOR'S MANUAL
EN
11. ERROR MESSAGES
Code
Message
Solution
USER COMMUNICATIONS
The shielded door connected to the
X-ray device is open
Close the door or any other device to
make X-ray emission safe for the
operator.
WARNING
0.1
X-ray Button released during
exposure
Hold the X-ray emission button until
the end of the procedure.
0.2
X-ray Button not released at the end
of examination
1) X-ray button was held for too
long after the end of exposure
(>15s).
2) A reset command was sent while
the xray button was being
pressed.
3) The device exited the stand-by
status while the xray button was
being pressed.
0.3
Emergency Button pressed
Release the emergency stop button.
0.4
POWER Circuit Board unable to
communicate with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
0.5
KEYBOARD Circuit Board unable to
communicate with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
0.6
CONVERTER Circuit Board unable
to communicate with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
0.8
Circuit Board of CEPH-Detector Lift
(CSENS) unable to communicate
with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
0.9
Circuit Board of Headrest Lift
(MSENS) unable to communicate
with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
0.10
Circuit Board of Detectors Carousel
(GSENS) unable to communicate
with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.
0.11
Circuit
Board
of
COLUMN-
INVERTER unable to communicate
with MAIN Board
Turn off the unit, check CAN bus
connections, wait 30sec and turn it
back on. If the problem persists
contact tech support.