Software Communication
3-25
T T
ERMINATE
C
OMMAND
The [T] command terminates plunger moves in progress ([A], [a], [P], [p], [D], and [d])
and delays [M]).
NOTE
The [T] command will not terminate Valve Move commands.
The [T] command will terminate both single commands and program strings. If a
program string is terminated before completion, the [R] (Execution) command will
resume the program string. If the command was terminated due to a problem or error, the
pump must be reinitialized.
When a plunger move is terminated, lost steps may result. Reinitialization is
recommended following termination.
NOTE
The [T] command is invalid in CAN communication. The equivalent
command is ASCII 4 for frame type 2.
J <n> A
UXILIARY
O
UTPUTS
The [J] command sets the TTL ouput lines.
The syntax for this command is:
[J<n>], where <n> = 0..7 (0 is the default)
The XP 3000 provides three TTL outputs on JP3 (pins 13, 14, and 15) that correspond to
outputs 1, 2, and 3. They are controlled as follows:
XP Command
Output 3 (Pin 15)
Output 2 (Pin 14)
Output 1 (Pin 13)
J0
0
0
0
J1
0
0
1
J2
0
1
0
J3
0
1
1
J4
1
0
0
J5
1
0
1
J6
1
1
0
J7
1
1
1
(0 = low; i.e., Gnd; 1 = high; i.e., +5VDC)
Содержание XP 3000
Страница 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Страница 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Страница 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Страница 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Страница 134: ......